Trend Data Access
7-16
Service Manual
create an alarm condition (for example, SENSOR DISCONNECT) and verify acti-
vation of the institution's nurse call system.
7.4.2
Setting Nurse Call RS-232 Polarity
The nurse call polarity can be set to a positive signal or a negative signal during
an alarm condition.
To set nurse call polarity
1.
While in normal monitoring mode, press MENU.
2.
Press CONNECTIVITY SETTINGS.
3.
Press NURSE CALL SETTINGS.
Figure 7-7.
Nurse Call Polarity Screen
4.
Press NORM + for a normally high OR press NORM - for a normally low setting.
5.
Press SAVE CHANGES to save the selected setting.
Содержание Nellcor GR101704
Страница 1: ...Service Manual Nellcor Bedside Respiratory Patient Monitoring System...
Страница 8: ...Page Left Intentionally Blank vi...
Страница 32: ...Page Left Intentionally Blank 2 14 Service Manual Product Specifications...
Страница 46: ...Page Left Intentionally Blank 3 14 Service Manual Theory of Operations...
Страница 54: ...Page Left Intentionally Blank 4 8 Service Manual Product Overview...
Страница 96: ...Operation 6 34 Service Manual a Press an alternate menu option b Press EXIT AND RESTART to exit SERVICE MODE...
Страница 114: ...Page Left Intentionally Blank 7 18 Service Manual Trend Data Access...
Страница 120: ...Page Left Intentionally Blank 8 6 Service Manual Performance Considerations...
Страница 168: ...Page Left Intentionally Blank 10 46 Service Manual Modification and Testing...
Страница 192: ...Repair 12 4 Service Manual Figure 12 2 Exploded View of Internal Components...
Страница 232: ...Page Left Intentionally Blank 12 44 Service Manual Repair...
Страница 235: ......