Nellcor™ Respiration Rate
ADD-30
Operator’s Manual Addendum
Figure 1-18.
Modified Bland-Altman Plot of Nellcor Respiration Rate Software versus Clinician Over-
scored Respiration Rate
In
Table 1-10
,
Local Bias with Respect to Reference Respiratory Rate
, the hos-
pital data is broken into sub-ranges of respiration rate. A performance analysis
is completed on each respiration rate sub-range. The limits of agreement were
derived from data in each respiration rate sub-range and are provided in
Table 1-10
.
1
Difference (Clinician Overscored Respiration
Rate - Nellcor Respiration Rate, BrPM)
3
Upper 95% LoA
4
Mean
2
Clinician Overscored Respiration Rate (BrPM)
5
Lower 95% LoA
Table 1-10.
Local Bias with Respect to Reference Respiratory Rate
Reference Respiration
Rate Range (Breaths/Min)
4-7
8-11
12-15
16-19
20-23
24-27
28-31
32
Mean Bias
-2.90
-1.82
-0.62
-0.04
0.50
0.49
2.28
N/A
Upper 95% LoA
1.99
3.70
3.26
3.86
7.33
3.27
7.14
N/A
Lower 95% LoA
-7.79
-7.35
-4.51
-3.95
-6.32
-2.29
-2.58
N/A