Nellcor™ Respiration Rate
ADD-14
Operator’s Manual Addendum
1.4.2
Required Pulse Oximetry Sensor Usage
To obtain respiration rate, caregivers must use a Nellcor™ Adult Respiratory
Sensor. The monitoring system reports the appropriate sensor type in the
message field when it detects the sensor.
Figure 1-9.
Sample Sensor Type Message
If caregivers use non-respiratory sensors, listed in the
Operator’s Manual
, the
monitoring system continues to post both SpO
2
and pulse rate data, with the
user message, NO RR SENSOR CONNECTED, and dashes in the RR field.
Figure 1-10.
RR field with No RR Sensor Message
If caregivers use a respiratory sensor, the monitoring system will post both
SpO
2
and pulse rate and the message CALCULATING until enough data are
available to post respiration rate. The monitoring system posts a status
message for instances where it is unable calculate a respiration rate. If such is
the case, examine all possible performance considerations. Reference
Figure 1-11.
RR field with Calculating Message