Specifications
A-4
Operator's Manual
A.2.6
Alarm Tone Definitions
A.2.7
Default Settings
Table A-6.
Alarm Tone Definitions
Tone category
Description
Medium priority alarm tone
Volume level
High, medium, and low
Sound pressure
57 dB (high volume), 49 dB (medium volume), 40 dB (low volume)
Pitch (±5 Hz)
720 Hz
Pulse width (±20 msec)
146 msec
Number of pulses in burst
3
Repetitions
Continually
Interburst interval
Approximately 8 sec
Low priority alarm tone
Volume level
High, medium, and low
Sound pressure
51 dB (high volume), 42 dB (medium volume), 34 dB (low volume)
Pitch (±5 Hz)
720 Hz
Pulse width (±20 msec)
187 msec
Number of pulses in burst
1
Repetitions
Continually every 16 seconds while visual alarm is presented
Interburst interval
Approximately 16 sec
Alarm reminder tone
Volume level
High, medium, and low
Sound pressure
50 dB (high volume), 42 dB (medium volume), 33 dB (low volume)
Pitch (±20 Hz)
720 Hz
Pulse width (±20 msec)
900 msec
Number of pulses
1
Repetitions
Continually every 3 minutes
Table A-7.
Default Settings
Default
Range
High alarm limit
OFF
20–95
Low alarm limit
40
15–90
Alarm volume
Medium
Low, medium, high
High rSO
2
scale limit
100
60–100 in increments of 10; 5 above high alarm limit
Low rSO
2
scale limit
30
0–30 in increments of 10; 5 below low alarm limit
Содержание INVOS 5100C
Страница 1: ...Operator s Manual INVOS Regional Saturation Patient Monitoring System Model 5100C...
Страница 12: ...Page Left Intentionally Blank x Operator s Manual...
Страница 26: ...Introduction Page Left Intentionally Blank 1 14 Operator s Manual...
Страница 30: ...Product Overview Page Left Intentionally Blank 2 4 Operator s Manual...
Страница 46: ...Installation Page Left Intentionally Blank 3 16 Operator s Manual...
Страница 50: ...Quick Setup Page Left Intentionally Blank 4 4 Operator s Manual...
Страница 142: ...Database Operational Mode and Area Under the Curve Page Left Intentionally Blank 8 24 Operator s Manual...
Страница 164: ...Specifications Page Left Intentionally Blank A 12 Operator s Manual...
Страница 170: ...Clinical Studies Page Left Intentionally Blank B 6 Operator s Manual...
Страница 176: ...Glossary Page Left Intentionally Blank C 6 Operator s Manual...
Страница 177: ......