![Covidien Force FX-C Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/covidien/force-fx-c/force-fx-c_user-manual_2673883090.webp)
Performance Characteristics
9-6
Force FX Electrosurgical Generator C User’s Guide
Expansion Port
The 15-pin connector supports the following signals:
–
Pin 2 – isolated transmit (serial data output transmit line)
–
Pin 3 – isolated receive (serial data input receive line)
–
Pin 5 – isolated ground (reference for transmit and receive)
–
Pin 9 – RF disable: input signal which, when activated by an external device,
disables active RF output
–
Pin 10 – RF current: output signal proportional to active RF current
–
Pin 11 – RF voltage: output signal proportional to active RF voltage
Expansion power (from the low voltage power supply):
+5 V (pin 6), – 12 V (pin 14), + 12 V (pin 15), and ground (pins 12 & 13)
Low Frequency (50–60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Enclosure source
current, ground open
< 300 μA
Source current,
patient leads, all
outputs
Normal polarity, intact ground: < 10 μA
Normal polarity, ground open: < 50 μA
Reverse polarity, ground open: < 50 μA
Sink current at high
line, all inputs
< 50 μA
Bipolar RF leakage
current
< 59.2 mA rms
Monopolar RF leakage
current
(additional tolerance)
< 150 mA rms
CEM output modes
< 150 mA rms at ð 50 W