Electrical Safety T

ests

ClosureRFG Radiofrequency Generator

11-7

Electromagnetic Compliance (EMC)

Conducted RF 
IEC 61000-4-6

3 V

RMS

150 kHz 

to 80

 

MHz

3 V

RMS

Portable and mobile RF 
communications equipment should be 
used no closer to any part of the 
ClosureRFG radiofrequency generator, 
including cables, than the 
recommended separation distance 
calculated from the equation applicable 
to the frequency of the transmitter.

Recommended separation distance

d=1.2 √P

d=1.2 √P   80 MHz to 800 MHz

d=2.3 √P   800 MHz to 2.5 GHz

where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation distance in 
metres (m).

Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey

a

 should be 

less than the compliance level in each 
frequency range.

b

Interference may occur in the vicinity of 
equipment marked with the following 
symbol:

Radiated RF IEC 
61000-4-3

3 V/m80 MHz 
to 2,5

 

GHz

3 V

RMS

NOTE U

T

 is the a.c. mains voltage prior to application of the test level.

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.

a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due 
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the RF generator is used exceeds the applicable RF 
compliance level above, the RF generator should be observed to verify normal operation. If 
abnormal performance is observed, additional measures may be necessary, such as re-orienting 
or relocating the RF generator.

b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Guidance and manufacturer’s declaration - electromagnetic immunity

Содержание ClosureRFG

Страница 1: ...Operation Manual ClosureRFG TM Radiofrequency Generator Covidien RFG3 Specs Provided by www AAATesters com...

Страница 2: ...under shall be limited to the aggregate purchase price for the Products sold to the customer This limited warranty is non transferable and runs only to the original purchaser of the covered Product s...

Страница 3: ...s internal business purpose in the operation of the Product or equipment purchased from other otherwise provided by COVIDIEN or its affiliates Except to the extent expressly authorized in this Softwar...

Страница 4: ...mer has not cured all breaches of this License s limitations or restrictions Upon such termination Customer will immediately pay all undisputed fees outstanding cease use of all Software return or del...

Страница 5: ...APPLY EVEN IF THE ABOVE STATED WARRANTY FAILS OF ITS ESSENTIAL PURPOSE SOME STATES DO NOT ALLOW LIMITATION OR EXCLUSION OF LIABILITY FOR CONSEQUENTIAL OR INCIDENTAL DAMAGES U S Government Rights The...

Страница 6: ...ClosureRFG Radiofrequency Generator vii Software License parties hereby submit to the jurisdiction of the courts of the Commonwealth of Massachusetts...

Страница 7: ......

Страница 8: ...ar Panel 1 5 Side Panels 1 6 General Safety Guidelines 1 6 Warnings and Safety Notices Setting Up the ClosureRFG Generator 1 7 Warnings and Safety Notices During RF Treatment 1 8 Unpacking and Inspect...

Страница 9: ...4 3 Starting RF Treatment 4 4 Stopping RF Treatment 4 5 Chapter 5 Treatment Times Retain Total Treatment Timer Connecting a New Catheter 5 1 Treatment Time Reset Using the Same Catheter 5 2 Export RF...

Страница 10: ...wer Output Charts 9 10 Chapter 10 Service Maintenance Maintenance 10 1 General Safety Guidelines 10 1 Fuse Replacement 10 2 Repairs and Returns 10 3 Return for Repair 10 3 Warranty 10 3 Cleaning and D...

Страница 11: ...s for use with Covidien ClosureFast endovenous radiofrequency catheters and Covidien ClosureRFS endovenous radiofrequency stylets Throughout this manual the term catheter generally refers to both comp...

Страница 12: ...as required for each geographical region The following RF treatment catheters Covidien ClosureRFS endovenous radiofrequency stylets with cord length of 8 ft Covidien ClosureFast 7 cm endovenous radio...

Страница 13: ...n be set to be compatible with the treatment environment Alarm medium priority Three rapid mid frequency mixed tones Informational Single short high frequency pure tone RFS Advisory Two short informat...

Страница 14: ...wing buttons and indicators AC Power Switch Illuminates when AC power is provided to the ClosureRFG generator RF Power Delivery Indicators Turns on when the ClosureRFG generator is first powered on an...

Страница 15: ...ock these openings AC Power Cord Wrap When the ClosureRFG generator is not in use unplug the unit from the AC power source and wrap the cord around this wrap area Top Handle Best way to lift and carry...

Страница 16: ...bration and repairs For qualified service personnel only Rubber Port Cover Provides a seal to limit liquid splashes getting into ports General Safety Guidelines The safe and effective use of RF energy...

Страница 17: ...r surgical drapes Electrosurgical instruments that are activated or hot from use can cause a fire When not in use place electrosurgical instruments in a safety holster or safely away from patients the...

Страница 18: ...eration is required during power interruptions the ClosureRFG generator must be powered from an uninterruptible power supply Warning Use the ClosureRFG generator only with Covidien ClosureFast endoven...

Страница 19: ...f electrosurgery could result in neuromuscular stimulation resulting in patient injury Electrodes and probes for monitoring stimulating and imaging devices can provide paths for high frequency current...

Страница 20: ......

Страница 21: ...e optimal use access all controls and view the operating screen Caution Do not stack any items on top of the ClosureRFG generator Doing so could damage the unit The ClosureRFG generator should not be...

Страница 22: ...t if any of its surfaces appear to be cracked or damaged Do not plug the ClosureRFG generator into a damaged outlet Do not use a damaged AC power cord Do not use extension cords multiple portable outl...

Страница 23: ...Version X X X Please Connect the Device Before inserting into body 1 2 The screen is divided into two areas with the data area above the message area The data area displays measurements temperature p...

Страница 24: ...Undo button Undo to undo any changes made since opening the Generator Setup screen Press the Home button to accept the new settings and exit this screen Setting System Language EN FR DE IT NL ES SV T...

Страница 25: ...30 31 29 30 25 26 27 28 2 To change the year press the left double arrow button to decrease or the right double arrow button to increase 3 To change the month press the left single arrow button to dec...

Страница 26: ...dd yyyy 7 Press the row that displays the preferred date format Setting System Time 11 50 AM 1 Press the hours or minutes button to change the corresponding value A keypad is displayed MM 7 8 9 4 5 6...

Страница 27: ...gs Time Adjusting Display Brightness 1 To change the brightness level Press the Brightness bar at the desired level or Press and drag the Slider button or Press the Dimmer or Brighter button 2 To retu...

Страница 28: ...ormational tone for all other parameter conditions sounds During RF treatment the gauge related to that parameter turns gray An alarm message is displayed indicating the problem The operator can press...

Страница 29: ...aligned with the red raised line on the catheter connector before inserting it Insert the connector end of a Covidien ClosureFast endovenous radiofrequency catheter or Covidien ClosureRFS endovenous...

Страница 30: ......

Страница 31: ...icates current temperature of the catheter it is not intended to measure patient body temperature Treatment Time Indicates by counting down the amount of time RF power has been delivered during the cu...

Страница 32: ...ges If a parameter is outside the advisory limits its gauge brightens within the red area or lights only the blue area a message is displayed and a tone is sounded ClosureFast gauges during treatment...

Страница 33: ...eFast catheter Press Device Handle Button to Treat ClosureFast Total 00 00 m s 0 RF Cycle s Parameters of the catheter are checked before RF treatment is allowed If the temperature and impedance are a...

Страница 34: ...or more information see Functional and Advisory Limits on page 2 8 Starting RF Treatment Note Review this section and General Safety Guidelines on page 10 1 before starting RF treatment Caution Do not...

Страница 35: ...gress ClosureFast Total 00 00 m s 0 RF Cycle s 0 10s 1 2 3 2 RF delivery stops automatically when the 20 second RF treatment cycle is complete Warning If RF treatment does not stop disconnect the cath...

Страница 36: ...Repeat step 3 until the procedure is completed Stopping RF Treatment RF treatment stops automatically since each cycle has a fixed time If it s found necessary to stop RF treatment before the timer fi...

Страница 37: ...tton Starts RF treatment Changes to the blue RF Stop button when pressed Stylet temperature Indicates current temperature of the connected stylet it is not intended to measure patient body temperature...

Страница 38: ...ges during treatment ClosureRFS 225 6 0w s 0 10 85 c 1 2 3 Illumination of the temperature gauge segments Blue Cool waiting for or at the beginning of a heating cycle below advisory limit during treat...

Страница 39: ...en RF Start button Ready to treat with a ClosureRFS stylet ClosureRFS 35 c 225 0 0w s 0 00 Press RF Button to Treat Parameters of the stylet are checked before RF treatment is allowed If the temperatu...

Страница 40: ...ghly read the stylet s instructions for use Confirm that the connected stylet matches the stylet type shown at the top of the Data Display area RF treatment can be stopped at any time if necessary by...

Страница 41: ...Stop button on the screen to stop RF treatment a The RF Stop button changes from blue to the green RF Start button b The Treatment Timer 3 in the previous illustration displays the time that RF power...

Страница 42: ...3 Move the stylet as directed in the stylet s instructions for use 4 Press the RF Stop button on the screen to pause RF treatment or when the procedure is finished Note The Treatment Timer displays th...

Страница 43: ...sed on either the same patient patient site or a new patient patient site The screen shows this question Would you like to retain the total treatment timer value 1 Would you like to retain the total t...

Страница 44: ...Timer Value button at the bottom of the screen The screen prompts for confirmation Would you like to retain the total treatment timer value 1 Would you like to retain the total treatment timer value...

Страница 45: ...ice should receive an explanation of the ESD warning symbol and training in ESD precautionary procedures This includes clinical biomedical engineering and health care staff At minimum this training sh...

Страница 46: ...st record is at the top of page 1 When entering this screen all entries not exported so far are already selected 4 Perform one of the following actions a If the selection is correct go to step 5 below...

Страница 47: ...2 14 2012 09 23 16 32 RFS 5 12 1 SD Cards The SD card stores ClosureRFG radiofrequency generator treatment data for troubleshooting these records are not saved in the ClosureRFG generator if the SD ca...

Страница 48: ......

Страница 49: ...one required High Power Advisory Display Advisory Power high Sound Informational Tone None required High Impedance Functional Advisory Display Advisory Impedance high Check Connection Sound Informatio...

Страница 50: ...Advisory Clear any advisory message Display Advisory Power high Sound Informational Tone None required Low Power And Low Impedance Advisory Clear any advisory message Display Advisory Power low Sound...

Страница 51: ...ledgement High Impedance Functional Limit Display Treatment Halted Impedance high Please press to continue Sound Informational Tone Wait for User Acknowledgement User acknowledgement Low Impedance Fun...

Страница 52: ...Y W Press Device Handle Button to Treat Sound Informational Tone None required Terminate ClosureFast Complete Cycle Press Device Handle button to Treat RF cycle completed at mm ss X C Y W Z Total Tre...

Страница 53: ...nds Output RF ramp up tone None required Confirm Discard Display Would you like to retain the total treatment timer value Wait for User Acknowledgement User acknowledgement Error Log Corruption Displa...

Страница 54: ...t Medium High Temperature Functional Limit Alarm Priority Sound Pressure Priority Sound Pressure Level Medium 45 dB Message Priority Priority Message 1 Highest Alarm Message 2 Advisory Message 3 Infor...

Страница 55: ...pected state Cycle system power and allow the POST testing to complete If the system does not respond and or the same error code is displayed contact Covidien customer service 2 Unexpected event 3 A p...

Страница 56: ...n on the About tab of the Settings menu If the serial numbers do not match and or the same error code is displayed contact Covidien customer service 28 The inserted USB device is not correctly formatt...

Страница 57: ...em power If the system does not respond and or the same error code is displayed contact Covidien customer service 37 The temperature of the ClosureRFG generator is above the specified operating temper...

Страница 58: ...insert Cycle system power and allow POST to complete If the issue persists reformat the USB device on your PC by right clicking the device and selecting format The device must be formatted to FAT32 If...

Страница 59: ...insert a known working catheter and cycle system power If the system does not respond and or the same error code is displayed contact Covidien customer service 102 Temperature is too high 103 Impedan...

Страница 60: ...110 Temperature and Power are too low 111 Temperature is too high and Power is too low 112 Impedance and Power are too low 113 Temperature Impedance and Power are too low 114 Temperature is too high...

Страница 61: ...oo low 205 Power is too high 206 Power is too low 314 Task took too long to complete Cycle system power and allow the POST to complete If the system does not respond and or the same error code is disp...

Страница 62: ...ID Calibration Load Cycle system power and allow the POST to complete If the error persists insert a known working catheter Cycle system power again If the system does not respond and or the same erro...

Страница 63: ...stem power cycle and allow the POST to complete If the system does not respond and or the same error code is displayed contact Covidien customer service 471 The ClosureRFG generator Calibration Load R...

Страница 64: ...has failed 481 The frequency source of the ClosureRFG generator is not correct 482 The ClosureRFG generator has too much offset in the catheter ID circuit Cycle system power and allow the POST to comp...

Страница 65: ...igns of physical damage Check the power cord Replace the power cord if you find exposed wires cracks frayed insulation or a damaged connector Verify that the wall outlet is supplying power Refer to Ch...

Страница 66: ...ry Temperature low Inadequate vessel exsanguination Verify that the catheter is properly positioned Check for flow and employ or improve compression techniques as necessary Saline in catheter connecto...

Страница 67: ...of saline If present contact Covidien Customer Service Caution Take care never to plug a wet catheter connector into the receptacle as this may damage the ClosureRFG generator Short circuit Replace ca...

Страница 68: ...n circuit Check stylet connection disconnect and then reconnect Replace stylet Power high Too much blood flow Pullback rate too fast Check and improve exsanguination Verify proper pullback rate Temper...

Страница 69: ...fter acknowledgement Recoverable errors may require the operator to respond to an error message or make a clinical decision Some errors require pressing the Green Check Mark button after which the Clo...

Страница 70: ......

Страница 71: ...external defibrillator while the RF catheter is in use Electrical ClosureRFG Generator Operates from an AC power source ranging from 100 V to 240 VAC 50 Hz to 60 Hz 5A Includes a DIN 42801 compatible...

Страница 72: ...nd 1060 hPa Conditions for Operating the ClosureRFG Generator 50 F 10 C 104 F 40 C Temperature between 10 C and 40 C 90 10 Relative humidity non condensing between 10 and 90 700 hPa 1060 hPa Ambient a...

Страница 73: ...s designed to limit the output control depending on the accessory being used For example ClosureFast endovenous radiofrequency ablation catheter 7 cm is limited to 40 W maximum ClosureFast endovenous...

Страница 74: ...ed time of RF delivery The ClosureRFG generator also interfaces with a sensor in the catheter to provide a continuous display of measured patient temperature prior to and during RF delivery The major...

Страница 75: ...e UI assembly It supplies 12 volts to the UI assembly and provides access to all the important signals coming from it The I O board provides the communication interface between the RF board and UI ass...

Страница 76: ...with air oxygen or nitrous oxide Do not operate the ClosureRFG generator in contact with flammable materials chemicals or substances The ClosureRFG radiofrequency generator should be used only with a...

Страница 77: ...r is not in use unplug it from the AC power outlet and wrap the cord around the AC power cord wrap area Do not carry or lift the ClosureRFG generator by the AC power cord or catheter cables The chemic...

Страница 78: ...Watts Power ClosureFast catheters RF power icon Treatment Status RF ready to start green Options button Opens the Settings screen ClosureFast catheters RF power icon Treatment Status RF is powered on...

Страница 79: ...ns for use Equipotentiality RX only The output is floating with respect to ground This side up Fuses Fragile Dangerous voltage Do not dispose in trash SD Card USB Device Storage device slots Explosion...

Страница 80: ...ndovenous radiofrequency ablation RFA catheter 7 cm Defibrillation proof type CF applied part 90 10 4 F 104 F 20 C 40 C to Humidity limitations operation 60 10 104 F 158 F 40 C 70 C to Humidity limita...

Страница 81: ...put Control Settings at 200 2 3 1 Power output watts Power output control settings watts ClosureFast endovenous radiofrequency ablation RFA catheter 3 cm Power Output vs Output Control Settings at 200...

Страница 82: ...Output Full Power Output Control Settings 2 3 1 4 5 Power output watts Impedance ohms ClosureFast endovenous radiofrequency ablation RFA catheter 7 cm ClosureFast endovenous radiofrequency ablation R...

Страница 83: ...General Safety Guidelines on page 1 6 In addition The AC power switch on the front panel serves as the means of disconnection from supply mains This switch also isolates the ClosureRFG generator elect...

Страница 84: ...osureRFG radiofrequency generator s AC power cord from the AC power inlet 1 Fuse Replacement 1 2 3 4 5 2 Insert a small blade screwdriver into the small slot located on the left side of the holder 2 a...

Страница 85: ...ien technician voids the warranty Cleaning and Disinfecting The ClosureRFG radiofrequency generator requires no scheduled maintenance other than cleaning external surfaces Cleaning should be performed...

Страница 86: ...0 25 Sodium Hypochlorite Disinfecting wipes or pads that contain only Sodium Hypochlorite are acceptable IPA in water 70 30 All surfaces including the bottom side of the ClosureRFG generator and any s...

Страница 87: ...edge of the applicable electrical safety standards which pertain to these tests Covidien assumes no responsibility for damage to the unit due to improper application of the following tests If in doubt...

Страница 88: ...s 240 V Parameter Conditions Passing Input Voltage Rated line 10 264 V Line Cord Resistance N A 100 m 25 A chassis cord gnd 10 seconds 100 m Earth Leakage Power Norm GND OPEN 500 A Earth Leakage Power...

Страница 89: ...med Test Cable to access patient connections Short the electrodes and thermocouple wires together for testing 4 For the 25 A test connect the ROD L leads to the chassis ground lug and to the ground pi...

Страница 90: ...agnetic compatibility EMC Part 4 2 Testing and measurement techniques Electrostatic discharge immunity test IEC 61000 4 3 2010 Electromagnetic compatibility EMC Part 4 3 Testing and measurement techni...

Страница 91: ...nvironment Emissions Test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 2 The ClosureRFG generator must emit electromagnetic energy in order to perform its intended funct...

Страница 92: ...ion in power occurs it may be necessary to reset the unit by disconnecting and reconnecting the catheter Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line...

Страница 93: ...3 V m80 MHz to 2 5 GHz 3 VRMS NOTE UT is the a c mains voltage prior to application of the test level NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not a...

Страница 94: ...wer not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of...

Страница 95: ...here are no internal adjustments to the RF power generator Covidien has created a Biomedical Department Inspection Procedure for the testing of the ClosureRFG radiofrequency generator This procedure o...

Страница 96: ...SU Analyzer the measurement tolerances of the ESU Analyzer for a given test must be added to the tolerances listed for the ClosureRFG generator Power Delivery Verification This test verifies the abili...

Страница 97: ...ce RF power time or a combination of these outside of which RF delivery will not start or stops immediately Impedance The effective resistance of the Covidien ClosureFast endovenous radiofrequency cat...

Страница 98: ...me and cycles to the current open record This is used when a catheter is reconnected and the same site is to be further treated Retain continues to add cycles and time to an open record RF Radiofreque...

Страница 99: ...their respective owner For use under U S Pat Nos 8 321 019 and 8 636 729 2014 Covidien Made in USA Printed in USA Covidien llc 15 Hampshire Street Mansfield MA 02048 USA Covidien Ireland Limited IDA B...

Отзывы: