Electrical Safety T
ests
ClosureRFG Radiofrequency Generator
11-5
Electromagnetic Compliance (EMC)
Electromagnetic Emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The ClosureRFG radiofrequency generator is intended for use in the electromagnetic environment
specified below. The customer or the user of the ClosureRFG generator should assure that it is
used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 2
The ClosureRFG generator must emit
electromagnetic energy in order to perform its
intended function. Nearby electronic equipment may
be affected.
RF emissions CISPR 11
Class A
The ClosureRFG generator is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Complies