Caution
Protect against harmful ingress of water or particulate matter. Generator classification: IPX0 (no
protection).
10-2
ClosureRFG Radiofrequency Generator
Fuse Replacement
Fuse Replacement
In the event of a fuse failure, replace the fuses using the following procedure, as shown:
Caution
Fuses must be replaced with 5 A/250 V, 5x20 mm fuses. Using other fuses might damage the unit.
1. Disconnect the ClosureRFG radiofrequency generator’s AC power cord from the AC
power inlet (1).
Fuse Replacement
1
2
3
4
5
2. Insert a small blade screwdriver into the small slot located on the left side of the holder
(2) and pry the retaining clip out a small distance to loosen the latch.
3. Insert the blade into the small slot located on the right side of the holder (3) and pry
the retaining clip out a small distance.
4. Pull the fuse holder drawer (4) out.
5. Replace the fuses (5).
6. Reinsert the fuse holder by pushing it until the holder’s front surface is flush with the
ClosureRFG generator surface and the latches engage.