Err
or Codes
ClosureRFG Radiofrequency Generator
7-9
468
More than 5 resets
occur within 1 minute.
Could occur when
operator repeatedly
cycles RF treatment.
Unplug the system’s power cord from the wall. Re-insert the
plug into the wall receptacle and allow the system POST to
complete. If the system does not respond and/or the same
error code is displayed, contact Covidien customer service.
469
The installed generator
software is not
compatible with the
HW revision.
Cycle system power and insert a USB device with the latest
system SW. Upgrade the system using the “About” tab.
Cycle system power again. If the system does not respond
and/or the same error code is displayed, contact Covidien
customer service.
470
The ClosureRFG
radiofrequency
generator Output
Relay is stuck closed.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
471
The ClosureRFG
generator Calibration
Load Relay is stuck
closed.
472
The ClosureRFG
generator Phase Offset
is too large.
473
The Safety Enable is
not functioning
correctly.
474
The ClosureRFG
generator Output RF
Indicator is fault.
475
The power supply is in
error more than
±10%.
476
The ClosureRFG
generator High Speed
ADC is saturate.
Cycle system power and insert a USB device with the latest
system SW. Upgrade the system using the “About” tab.
Cycle system power again. If the system does not respond
and/or the same error code is displayed, contact Covidien
customer service.
477
The ClosureRFG
generator HW Power
Limit is faulty.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
Error
Code
Description
Corrective Action