7-8
ClosureRFG Radiofrequency Generator
360
The HW Power limit
was exceeded.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
363
The output relay was
open when it should
have been closed.
364
A broken
thermocouple was
detected.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter. Cycle
system power again. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
450
Generator ROM
Failure.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
451
Generator RAM
Failure.
452
Generator Watchdog
Failure
455
Generator power
supply temperature
exceeds specification.
457
The Catheter ID
circuitry failed to
correctly read the ID
Calibration Load.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter. Cycle
system power again. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
459
The calibration
coefficients are
corrupt.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
460
Generator failed
calibration check.
461
The Primary Side
Power Measurement
and Catheter Side
Power Measurement
differ by more than the
allowed value.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter. Cycle
system power again. If the system does not respond and/or
the same error code is displayed, contact Covidien customer
service.
462
The Isothermal
Reference and
Isothermal Backup
differ by more than the
allowed value.
Cycle system power cycle and allow the POST to complete.
If the system does not respond and/or the same error code is
displayed, contact Covidien customer service.
Error
Code
Description
Corrective Action