Err
or Codes
ClosureRFG Radiofrequency Generator
7-5
101
Temperature is too
low.
Cycle system power and allow the POST to complete. If the
error persists, insert a known, working catheter and cycle
system power. If the system does not respond and/or the
same error code is displayed, contact Covidien customer
service.
102
Temperature is too
high.
103
Impedance is too low.
104
Temperature and
Impedance are too
low.
105
Temperature is too
high and Impedance is
too low.
Error
Code
Description
Corrective Action