3-2
ClosureRFG Radiofrequency Generator
Ready-to-Treat Screen
ུ
RF Power:
Indicates RF power actively applied to the catheter.
ཱུ
Treatment Summary:
Indicates the total time and number of treatment cycles since
reset, power on, or change of catheter, as minutes:seconds.
ClosureFast Catheter Gauges
Data Display with Gauges
Gauges indicate the measured parameter and whether the catheter’s parameters
(temperature and RF power) are within acceptable ranges. If a parameter is outside the
advisory limits, its gauge brightens within the red area or lights only the blue area, a
message is displayed, and a tone is sounded.
ClosureFast gauges during treatment
ClosureFast
Total
00:00 m:s 0 RF Cycle(s)
0:10
s
1
2
The illumination of the
temperature
gauge segments:
•
Blue:
Cool; waiting for, or at the beginning of a heating cycle; or low during
treatment
•
Blue-Green:
Within normal range after initial heating
•
Red:
Temperature above advisory or functional limit
ཱ
The illumination of the
power
gauge segments:
•
Blue:
Low but within limits
•
Blue-Green:
Within normal range
•
Red:
Power above advisory or functional limit