background image

 

- 9 - 

4) The gas sensitivity can be set from five levels in descending order of sensitivity 0, 

1/5, 1, 2 and 3. Change the gas sensitivity as needed when using the Gas Leak 
Detector. 

 
 
 
 
 
 
 
 
 

  Decreasing the Gas Sensitivity Level 

Each  time  pressing  the  power  button,  the 
buzzer  will  sound  once  and  the  gas  sensitivity 
level  will  be  decreased  one  level.  The  lighted 
gas  sensitivity  lamp  (green)  will  move  up  one 
by one, and the LCD will also display the same 
sensitivity level as shown on the gas sensitivity 
lamp. 
When  the  gas  sensitivity  level  is  decreased 
beyond 3, the level will return to level 0. 

 
 
 
 

 

 
 
 

 

  Increasing the Gas Sensitivity Level 

Each time pressing the power button twice (like 
double-click),  the  buzzer  will  sound  twice  and 
the  gas  sensitivity  level  will  be  increased  one 
level.  The  lighted  gas  sensitivity  lamp  (green) 
will  move  down  one  by  one,  and  the  LCD  will 
also display the same sensitivity level as shown 
on the gas sensitivity lamp. 
When  the  gas  sensitivity  level  is  increased 
beyond 0, the level will return to level 3. 

 
 

 
 

 

  The  detected  gas  concentrations  for  the  gas  sensitivity  levels  are  as 

follows. 

Level 0: Detects concentration of 10 ppm or less. 

(Detectable gas leakage: 3.3 × 10 

-6

 Pa·m

3

/s) 

Level 1/5: Detects concentrations of around 30 ppm. 

  The  gas  sensitivity  level  is  always  1/5  for  Freon  gas  and  cannot  be 

changed. 

NOTE 

Sensitivity setting:

 

 

 

3

 

 

  2 

 

 

1

 

 

 

1/5

 

 

 

 

Press the power button for more than 3 seconds to turn OFF the power. 

NOTE 

 

Sensitivity setting:

 

 

 

  3

 

      2 

 

 

1

 

 

 

1/5

 

 

 

LOW 

Gas sensitivity

SENS. 

SENS. 

Gas sensitivity 

HIGH 

Содержание XP-702III-A(S)

Страница 1: ...xplains the use of three models listed on the left Keep this instruction manual available for quick reference when needed Read this instruction manual and throughly understand before using the product...

Страница 2: ...get Gas 13 Silent Mode No Operation Sound 13 Deactivating the Alarm Buzzer 14 Lighting the Backlight 14 4 Error Displays Error Alarms 15 5 Replacement of Consumable Parts 16 Filter Element Replacement...

Страница 3: ...tely 1 Refer to page 20 for details on the contents Name Quantity Gas Leak Detector with soft case 1 DF 114 Drain filter 1 Replacement FE 2 filter element 1 AT 2G Gas Probe 1 FE 106 Dust filter 1 Hand...

Страница 4: ...nual Precautions for Explosion proof Applications Observe the following precautions in any explosion proof application Symbols The following symbols are used for safety purposes Indicates an imminentl...

Страница 5: ...ies if the batteries are leaking when the remaining battery charge is low The Gas Leak Detector has an explosion proof structure To avoid damaging the explosion proof performance never attempt to disa...

Страница 6: ...4 2 Part Names and Functions 1 2 11 7 9 5 4 10 8 6 3 13 12...

Страница 7: ...is very high These lamps also flash when an error occurs in the pump or sensor 9 Buzzer hole The buzzer sounds 10 LCD Displays the gas sensitivity and messages a The remaining battery charge is displa...

Страница 8: ...6 No Name Functions 1 Side Zipper Open the zipper to replace the batteries 2 Back Pocket The Operating Guide is stored here 3 Inside Pocket A Replacement Filter Element is stored here Soft case 1 3 2...

Страница 9: ...warm up is completed the highest sensitivity level 0 lamp will flash The maximum warm up is three minutes Always perform the daily check before using the Gas Leak Detector Refer to page 19 WARNING ST...

Страница 10: ...gas sensitivity level SENS L H Check the target gas before detection For a model of one gas type XP 702III B S check the gas name with gas name label on the Gas Leak Detector For a model of two gas ty...

Страница 11: ...twice like double click the buzzer will sound twice and the gas sensitivity level will be increased one level The lighted gas sensitivity lamp green will move down one by one and the LCD will also dis...

Страница 12: ...remaining battery charge indicator appears replace the batteries in a non hazardous clean air location promptly Refer to Battery Replacement on page 18 If E B is displayed on the LCD refer to Battery...

Страница 13: ...a high concentration gas of any in the following table is detected leave the Gas Leak Detector in clean air for a certain period of time before turning OFF the power Turning OFF the power immediately...

Страница 14: ...restart pump button To display the gas sensitivity level again press the restart pump button while the gas name is being displayed By pressing the power button while the gas name is displayed the gas...

Страница 15: ...mode press the buzzer button for 3 seconds The alarm buzzer cannot be turned ON or OFF unless silent mode is released Refer to Deactivating the Alarm Buzzer on page 14 The name of the target gas will...

Страница 16: ...utton is being pressed It also lights automatically for approximately 15 seconds when an error occurs When the power is turned ON again the alarm buzzer sound setting will be the same as before the po...

Страница 17: ...r turning ON the power several times there may be a problem with the sensor Request repairs Pump Error Error lamps flash Water or liquid may have been sampled or the inlet may be blocked Remove the wa...

Страница 18: ...unction properly and will not be able to detect correctly Attach the filter case securely If the filter case is loose water or other foreign matter may leak into the Gas Leak Detector If water or liqu...

Страница 19: ...Turn the cap of the Drain Filter counterclockwise to remove together with the Gas Probe 2 Remove FE 106 Dust Filter and replace with a new filter 3 Attach the Drain Filter and reassemble Make sure to...

Страница 20: ...Leak Detector and install two new batteries size AA alkaline as shown on the bottom of the compartment 3 Close the battery lid firmly 4 Zip the right and left of the soft case and close the Velcro ta...

Страница 21: ...the alarm sound indicator is being displayed and the alarm lamp red does not flash or the alarm buzzer does not sound contact your New Cosmos distributor Gas Probe Check the Gas Probe for any wear or...

Страница 22: ...Collector AT 7A 20524432 Used together with RP 4 Sampling Hose SH 4 1 20524429 Used together with Annealed Copper Tube and RP 1 Sampling Hose SH 4 05 20524531 Used together with Annealed Copper Tube...

Страница 23: ...attery polarity is reversed Insert the batteries correctly The power will not turn ON after pressing the power button Battery exhaustion Replace the batteries Battery Replacement on page 18 Displaying...

Страница 24: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation...

Страница 25: ...x ia II BT3 intrinsically safety Degree of Protection 3 Equivalent to IP22 4 Operating Temperature and Humidity range 20 to 50 degrees C less than 85 RH with no condensation Dimensions Approx 38 W x 1...

Страница 26: ...f Structure A structure applicable to an electrical device that prevents the electrical device from acting as a source of ignition and igniting explosive gas in the vicinity Intrinsically Safety A str...

Страница 27: ...MEMO...

Страница 28: ...Distributor Manufacturer New Cosmos Electric Co Ltd 2 5 4 Mitsuya naka Yodogawa ku Osaka 532 0036 Japan Phone 81 6 6309 1505 Fax 81 6 6309 1514 e info new cosmos co jp http www new cosmos co jp...

Отзывы: