Step 2 Inspection. 

Inspect the items for damage at all stages of handling. If damage is detected on any of the components it should be identified 

and documented. Complete the disinfection/sterilization process and contact technical service for replacement. 

Step 3 Presoak. 

Protective attire is required of personnel handling contaminated items. At the processing area soak or rinse the items in tap water 

22°-43° C. Please note that rinse with flowing water is not possible on the turbine. If an enzyme product is required, soak for one to two minutes. 

Remove and examine, extend the soak time for components with dried-on matter, prolonged soaking of components may be detrimental, causing 

damage to the component surfaces. Refer to the detergent instructions for its usage and soak time. 

Step 4 Disassembly. 

Disassemble the item (if necessary) according to the instructions reported in the corresponding section. 

Step 5 Cleaning. 

Protective attire is required for personnel handling contaminated items. Manual cleaning must be done in a manner that 

protects personnel handling the devices from aerosolization and splashing of infectious material. 
1.

 

Manual cleaning of the items should be done under 22°-43°C water. Use a neutral pH (7) mild detergent. Water hardness, temperature and 

the type of soil affect the effectiveness of the detergents; the detergent manufacturer's instructions should be consulted. Use a small soft 
brush to scrub all parts. Abrasive cleaning compounds and implements can damage the items and should not be used. Additional cleaning 
supplies may be required to clean stubborn stains or hard-to-reach areas. 

2.

 

Items must be thoroughly rinsed with clean water to remove the detergent residuals and debris from the components. Use a flowing triple 
rinse cycle at a minimum with tap water. Please note that rinse with flowing water is not possible on the turbine. 

3.

 

Dry all components thoroughly using soft clean clothes or disposable paper towels. 

 

Disinfection 

The recommendations in this sections have been retrieved from: 
William A. Rutala, Ph.D., M.P.H., David J. Weber, M.D., M.P.H., and the Healthcare Infection Control Practices Advisory Committee  (HICPAC): 

Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008 
(http://www.cdc.gov/hicpac/pdf/guidelines/Disinfection_Nov_2008.pdf) 
High-level disinfection is the recommended decontamination procedure for semi-critical devices. 
Devices are classified semi-critical when they touches mucous membranes or broken skin. Examples of semi-critical devices are flexible endoscopes, 

laryngoscopes, endotracheal tubes, respiratory therapy and anesthesia equipment, diaphragm fitting rings, and other similar devices. 

Preparing the disinfecting solution 

The recommended disinfection solutions are as follows: 

 

Sodium hypochlorite 0.5% (5000 ppm) prepared fresh for use within 24 hours. 

 

Sodium hypochlorite 1% (10000 ppm) prepared fresh for use within 30 days. 

The first solution can be prepared by adding 1 part household bleach (sodium hypochlorite 5.25%) to 9 parts water. The second solution can be 

prepared by adding 1 part household bleach to 4 parts water. 

 

. The turbine flowmeter 

Guidelines recommend that the turbine should be cleaned and disinfected prior to every test to ensure accurate measurements and to comply with 
recommended sanitation measures as follow. 

Cleaning the turbine 

Follow the standard cleaning procedure reported above, paying attention to the following: 
1.

 

For rinsing, do not use flowing water, which may damage the turbine. Rinse the turbine in a vessel, filled of clean water, shaking gently to 

remove the disinfectant. Do not place the turbine under running water or move the turbine while submerged. 

2.

 

For cleaning and rinsing, do not wet the sampling line. 

3.

 

Use the brush (point 1 step 5) only for the external parts of the turbine, in order to avoid damages to the turbine blade. 

Disinfecting the turbine 

1.

 

Take out the turbine. 

2.

 

Dip it in a disinfectant solution (non alcoholic based) for about 20 minutes. 

3.

 

Rinse the turbine in a vessel, filled of clean water, shaking gently to remove the disinfectant (do not clean the turbine by putting it under 

running water!). 

 

System maintenance - Cleaning and disinfecting - 27 

Содержание microQuark

Страница 1: ...Manuale Utente User Manual Manuel d utilisation Benutzerhandbuch Manual del Usuario microQuark PC based Spirometer...

Страница 2: ......

Страница 3: ...ransmitted inany form withoutthe expressed writtenpermissionof COSMED Srl Eachcopy of theCOSMED Softwarecanonly beinstalled ononecomputer Excel is aregistered trademark of MicrosoftCorporation DBIIIis...

Страница 4: ...2...

Страница 5: ...e 9 EMC 10 Overview of the manual 14 Introduction 15 MicroQuark overview 16 MicroQuark unit 16 Theflowmeter 16 Paper mouthpieces 17 Antibacterial filters 17 Noseclips 17 Installation 19 Beforestarting...

Страница 6: ...razionedi conformit 32 Service Warranty 33 Warranty and limitationof liability 33 Returngoods policy for warranty or nonwarranty repair 33 Repair ServicePolicy 33 Privacy Information 35 Personal datat...

Страница 7: ...Getting started...

Страница 8: ...nd trained personnel Environmental and electrical conditions in which the device operates should be in compliance with the specifications of this manual In particular grounding reliability and leakage...

Страница 9: ...equipment 14 The device should not be used adjacent to or stacked with other equipment If this is necessary you must verify that the device continues to operatenormally intheconfigurationinwhichit wil...

Страница 10: ...sts Hyperventilationsyndrome Contraindications for Bronchial Provocation testing Bronchial Provocation testing must be executed under the direction of a physician Testing is considered safe when execu...

Страница 11: ...d be operated under normal environmental temperatures and conditions which are defined as follows IEC 60601 1 EN 60601 1 Temperatures range 10 C 50 F and 40 C 104 F Relativehumidity range 30 to 90 not...

Страница 12: ...The device uses RF energy only for its internal function Therefore its RF emissions are very low and arenotlikely tocauseany interferencein nearby electronic equipment RF emissions CISPR 11 Class B T...

Страница 13: ...common mode 1kV differential mode 2kV commonmode Mains power quality should be that of a typical commercial or hospital environment Voltagedips short interruptions and voltage variations onpower suppl...

Страница 14: ...mitters as determined by an electromagnetic site surveya should be less than the compliancelevel ineachfrequency rangeb Interference may occur in the vicinity of equipment marked withthe following sym...

Страница 15: ...cy oftransmitter m 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 70 3 70 7 38 100 11 70 11 70 23 33 For...

Страница 16: ...equiredtoproperly install thedevice Systemmaintenance Describes system maintenanceprocedures Appendix Contains information regarding the warranty treatment of personal data reference standards technic...

Страница 17: ...intelligent flowmeter that connected to a Personal Computer laptop or desktop turns itinacompletespirometric lab The system is composed by the turbine flowmeter the measurement and data elaboration d...

Страница 18: ...The air passing through the helical conveyors takes a spiral motion which causes the rotation of the turbine rotor The rolling blade interrupts the infrared light beamed by the two diodes of the read...

Страница 19: ...infection control However regular cleaning and decontamination of lung function equipment should always beperformed Note The use of antibacterial filters is recommended even when using disposable mout...

Страница 20: ......

Страница 21: ...Installation...

Страница 22: ...should assure that the package contains the items listed in the packing list enclosed with the device If there are any missing or damaged parts you should contactCosmed s technical assistance Options...

Страница 23: ...USBplug Theunitis turnedon off automatically by thePC Insertthe turbineand besureto pushthe turbineup totouchtheend of theholder MicroQuark connects tothePCas showninthefollowing picture 1 microQuark...

Страница 24: ...Piani di Monte Savello 37 00041 AlbanoLaziale Rome ITALY Voice 39 06 931 5492 Fax 39 06 931 4580 email customersupport cosmed it Internet http www cosmed com Complaints feedback and suggestions If yo...

Страница 25: ...System maintenance...

Страница 26: ...oQuark arenon toxic and posenosafety risk tothepatientor operator Prior tothedevicecleaning disinfectionand inspectionit is necessary to switchoff the deviceand todisconnectadapters from thesupply mai...

Страница 27: ...oplet nuclei are likely to be encountered proper attention to environmental engineering controls suchas ventilation air filtration or ultravioletdecontaminationof air should beused topreventdiseasetra...

Страница 28: ...or partial disassembly for cleaning and disinfection It is the responsibility of the user health care personnel for ensuring that the cleaning methods recommended can be duplicated in their environmen...

Страница 29: ...dations inthis sections havebeen retrieved from William A Rutala Ph D M P H David J Weber M D M P H and the Healthcare Infection Control Practices Advisory Committee HICPAC Guidelinefor Disinfectionan...

Страница 30: ...epropeller rotates freely evenwith alow speed air flow 6 Connect the turbineto thereader Precautions to take when cleaning disinfecting and drying the turbine Donot exposethe turbineto highheat or to...

Страница 31: ...of cables plugs and any other visible part by means of simpleinspection whentheequipmentis switched off and adapters or electrical feeders aredisconnected from thesupply mains Extract the turbine from...

Страница 32: ......

Страница 33: ...Appendix...

Страница 34: ...ts of theAnnex I of theMedical DeviceDirective93 42 EEC areclassified inClass IIa their design manufacturing and final checks are performed according the Cosmed s Quality System conform to ISO 9001 20...

Страница 35: ...navoid of the warranty The limited warranty shall not be enlarged diminished or modified by the renderings of technical service from COSMED s agents or employees whentheproductis ordered or following...

Страница 36: ...39 06 9314580 E mail customersupport cosmed it USA contact COSMED USA Inc 2211 NorthElston Suite305 ChicagoIL 60614USA Phone 1 773 645 8113 Fax 1 773 645 8116 email usa sales cosmed it Toensurethatyou...

Страница 37: ...l invoices e Toprovideinformationtotheselected business partners needed tosupply your service How your personal data is treated Your personal data will be stored in an electronic format and protected...

Страница 38: ...ent must becollected separately according to the European Directive 2002 96 EEC Otherwiseitcancausedangerous consequences for theenvironmentand human health Thecrossed outwheeled binmeans thattheprodu...

Страница 39: ...device Devicenotsuitableinthepresenceof flammableanaesthetics Continuous functioning equipment EMC Thesystem meets the Standard IEC 60601 1 2 Quality Assurance UNIEN ISO 9001 2008 Registration n 387 A...

Страница 40: ...Flowmeter Bidirectional digital turbine Flow Range 0 16 l s VolumeRange 12l Accuracy 2 or 20ml s Resistance 0 6cmH2O l s 14l s Mouthpieces 31 and 22 mm Dimensions 150x 45 x 53mm Weight 77g 38 Appendix...

Страница 41: ...Standard for healthy non smoking adults ARRD Vol 10 3 p 57 67 1971 Pereira Brazil Pereira CAC Barreto SP Sim es JG Pereira FWL Gerstler JG Nakatani J Valores de Refer ncia para Espirometria em uma amo...

Страница 42: ...pressures in adults normal values and their relationship to determinants of respiratory function Bull Eur Physiopathol Resp 23 435 439 Automatic diagnosis algorithm Reference Lung FunctionTesting sele...

Страница 43: ...eadeeper breath dFVC 200ml and 5 bestFVC Blow outfaster dFEV1 200ml and 5 FEV1 Thatwas a good test Noerrors FVCreproducible Diff 2maxFVCwithin 0 2l FEV1reproducible Diff 2maxFEV1within0 2 l PEF reprod...

Страница 44: ...D Stanescu D Rodenstein C Veriter Reassesment of interrupter technique for measuring flow resistance in human J Appl Physiol 67 3 933 937 1989 Oximeter National Lung HealthEducationProgram NLHEP Guide...

Страница 45: ......

Страница 46: ...www cosmed com COSMEDsrl ViadeiPianidiMonteSavello37 AlbanoLaziale Rome 00041ITALY O 3906931 5492 F 3906931 4580 info cosmed com cosmed com...

Отзывы: