12 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 13

10. Pipettor Maintenance 

Depending on the applications and the intensity of use, 

the pipettor requires periodic maintenance. 

The components most exposed to vapors of aggressive 

solutions, such as shaft elements, should be regularly 

checked and cleaned.
To properly carry out the maintenance, familiarize yourself 

with the design of the shaft holder set shown in Figures 

10A, 10B, 10C, and follow the instructions given in this 

Section. 

 Do not use sharp tools for pipettor maintenance. It may 

cause damage to the device and affect the user’s safety.

Cleaning

External surfaces of the pipetting pushbutton, the ejector

 

pushbutton, the handgrip, the shaft nut, and the adjustment 

knob may be cleaned using a cloth dampened in isopropyl 

alcohol. The remaining parts removed from the pipettor 

during pipettor disassembly may be washed with distilled 

water or isopropyl alcohol.

NOTE:

 Before using cleaning agents other than those 

recommended by the manufacturer, check the compatibility 

charts and consider chemical resistance of the following 

plastics: PP, PC, POM, PA, PPS, PVDF used to produce the 

pipettor parts.
Dismantling shaft elements of the single-channel pipettor:
•  Holding the pipettor handle, slightly unscrew the shaft 

holder set and pull the Ejector (H) out of it;

•  Still holding the handle, unscrew the remaining part 

of the shaft holder set from it.

Then proceed according to the pipettor models shown 

below:

See ISO 8655 for the full table of the value of the conversion 

factor Z.

NOTE:

 Pipetting should be performed in accordance with 

the rules described in Sections 7 and 8.

Checking the Pipetting Accuracy Parameters

•  Set the pipettor to the calibration volume.
•  Perform 10 aspirations, and calculate their average value 

in [mg].

•  Calculate the volume in [μL] by multiplying the value 

in [mg] by the conversion factor Z [μL/mg].

If the average aspirated volume exceeds the permissible 

values of the range given in Section 5, the pipettor should 

be recalibrated.

Pipettor Recalibration

Make sure that the pipettor is set to the calibration volume 

and the volume adjustment knob is locked.

1. Remove the cap from the ejector pushbutton.
2. Set the calibration switch to the upper 

position (CAL) using the flat end of the 

calibration key.

3. Insert the calibration key with the cross 

end vertically into the hole in the ejector 

pushbutton, placing it in the counter’s 

calibration knob.

4. Turn the key so that the volume indicated 

by the counter is equal to the volume 

calculated.

5. Remove the calibration key.
6. Set the calibration switch in the 

operating position.

7. Fit the ejector pushbutton cap.
8. Repeat the pipetting accuracy check (Section 9: Checking 

the pipetting accuracy parameter). If the accuracy error 

value is exceeded, repeat the calibration process.

“Operating”

position

“Calibration”

position

calibration key

Содержание Lambda EliteTouch

Страница 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Страница 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Страница 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Страница 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Страница 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Страница 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Страница 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Страница 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Страница 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Страница 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Страница 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Страница 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Страница 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Страница 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Страница 16: ......

Страница 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Отзывы: