10 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 11

the preset volume) exceeds the permissible value given 

in the table in section 5, the pipettor recalibration procedure 

should be carried out.
Recalibration of the pipettor consists in setting on the counter 

the value of the liquid volume obtained in weighing. 

Recalibration of the pipettor is performed only for one liquid 

volume drawn by the pipettor. The recommended volume 

for recalibration is 10% of the maximum (nominal) value.

Conditions for Checking the Pipetting Accuracy Parameters

The pipetting accuracy is influenced by such factors as: 

tips used, characteristics of the pipetted liquid (density, 

viscosity), or operating conditions (ambient temperature, 

pressure).
To determine the accuracy error of the pipettor, the following 

conditions should be met:
•  Ambient temperature and the temperature of the 

pipettor, pipet tips, and liquid should be within the range 

of 20°C to 25°C and stabilized during weighing within 

±0.5°C.

•  Measurements should be conducted using distilled water.
• 

Balance sensitivity should be suitable for the volume “V” 

to be measured:

Volume Checked

(μL)

Balance Sensitivity

(mg)

0.1≤

V

≤ 10

0.001

10≤

V

≤100

0.01

V

>100

0.1

•  When calculating the liquid volume aspirated by 

the pipettor, the values of conversion factor Z [μL/mg] for 

distilled water or a liquid with comparable density should 

be taken into account. Sample values of this conversion 

factor are given in the following table: 

Temperature

(°C)

Pressure (kPa)

95.0

101.3

105.0

20

1.0028

1.0029

1.0029

21

1.0030

1.0031

1.0031

22

1.0032

1.0033

1.0033

23

1.0034

1.0035

1.0036

24

1.0037

1.0038

1.0038

25

1.0039

1.0040

1.0040

Aspirating High-density Liquids 

When pipetting liquids of higher viscosity or lower surface 

tension than water (e.g., sera or organic solvents), a film of 

liquid is formed on the inside wall of the pipet tip. This film 

can create an error. As the film remains relatively constant in 

successive pipetting operations with the same tip, this error 

can be eliminated by forming the film before transferring 

the first sample. This is done by aspirating a sample and 

dispensing it back into the same vessel. Now, as the film is 

already formed, all of the following samples will have better 

accuracy and repeatability.
This pre-rinsing operation should be repeated when 

the volume to be aspirated is changed or when a new pipet 

tip is used.

NOTE:

 Normally the degree of error resulting from heavy or 

viscous liquids is negligible if pipetting is done slowly and 

carefully. It is most important to give the liquids some time 

to react to the change of pressure by holding the pipet tip 

in its position for at least 2 seconds after the aspiration and 

the blow out stroke.

If in extreme cases, this method of operation does not result 

in accurate values, recalibrate the pipettor out in accordance 

with Section 9.
After calibration, it is recommended to record the recalibration 

and the size of the correction introduced, to facilitate 

reverse calibration to the standard liquid.

9. Checking Pipetting Accuracy Parameters 

     and Pipettor Recalibration

The pipettors are factory-calibrated using the gravimetric 

method with Corning® pipet tips and distilled water, in 

accordance with ISO 8655 guidelines for the maximum 

(nominal) liquid volume drawn by the pipettor and for 10% 

of the maximum volume according to the values given 

(Section 5).
The design of the pipettors enable their recalibration 

and adaptation to other pipetting techniques or liquid 

properties (i.e., temperature, density and viscosity).
It is recommended to carry out periodic checks of the 

operation of the pipettor at least once a year, the period 

of operation of the pipettor with no checks depends 

on the type of workload (liquids collected), and other 

conditions (load intensity and sterilization in the autoclave, 

replacement of components) in which the pipettor is used.

If during pipettor operation you find that the accuracy error 

(the difference between the real aspirated volume and 

Содержание Lambda EliteTouch

Страница 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Страница 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Страница 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Страница 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Страница 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Страница 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Страница 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Страница 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Страница 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Страница 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Страница 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Страница 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Страница 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Страница 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Страница 16: ......

Страница 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Отзывы: