20 – Corning Lambda EliteTouch Pipettors 

INSTRUCTION MANUAL – 21

Problem

Cause

Solution

Droplets 

of air appear 

in the liquid 

aspirated 

into the pipet 

tip.

The pipet tip immersion 

is too shallow.

Immerse the tip 

deeper according to 

the instructions.

The pipet tip is incorrectly 

pressed onto the pipettor 

shaft.

Press the pipettor 

firmly.

The tip is damaged or worn 

out due to extensive use.

Replace the tip with 

a new one.

The pipettor 

incorrectly 

aspirates 

the liquid 

or the liquid 

drops out 

from the tip.

The pipettor tip is 

incorrectly pressed onto 

the pipettor shaft.

Press the pipet tip 

firmly.

The shaft surface is 

damaged or contaminated 

at the sealing site.

Clean the shaft or 

replace it with a new 

one.

The plunger or the seal is 

damaged due to prolonged 

aspiration of aggressive 

liquids.

Disassemble the 

shaft set; wash the 

shaft, the plunger and 

the seal (Section 10: 

Cleaning). Replace the 

elements with new 

ones if necessary.
Apply a small amount 

of lubricant onto 

the plunger and 

reassemble the set 

in the correct order.

The inside of the pipettor is 

contaminated.

No lubrication on the 

sealing elements.

Uneven 

work of the 

pipetting set, 

the pipetting 

pushbutton 

gets blocked.

The inside of the pipettor 

is contaminated due to 

aspiration of aggressive 

substances.

Unscrew the shaft 

set, wash the parts. 

Replace the elements 

with new ones if 

necessary. Apply 

a small amount 

of lubricant on 

the plunger and 

reassemble the set 

in the correct order.

The inside of the pipettor 

is contaminated due to the 

liquid entering the pipettor.

No lubrication on the 

sealing elements, 

e.g. after repeated 

autoclaving procedures.

Sterilization 
Sterilization in an autoclave

The pipettor can be sterilized in an autoclave at 121°C for 

20 minutes. Sterilization under other conditions may cause 

damage to the pipettor. It is recommended to:
•  Unscrew the shaft nut slightly prior to sterilization and 

re-tighten after autoclaving. 

•  Set the locking ring in upper (unlocked) position prior 

to sterilization. 

•  Sterilize the pipettors in autoclave with an initial vacuum 

and drying cycle.

•  After sterilization, the pipettor should be dried and cooled 

to room temperature.

The precision of the results should not alter if the pipetting 

process and autoclaving are carried out as described in this 

manual. Because a slight change in the accuracy of the 

dosage may occur, it is recommended to:
• 

Check the calibration of the pipettor after the initial first, 

third, and fifth autoclaving cycles and then after every 

10 autoclaving cycles.

Ultra Violet (UV) Sterilization

The pipettors are UV resistant. However, the recommended 

distance from the radiation source to the exposed element 

should be not less than 50 cm. Prolonged and very intense 

UV exposure can cause discoloration of pipettor parts 

without effecting their performance.

11. Troubleshooting 

If you notice an improper pipettor operation, identify 

the cause and eliminate the fault. Follow the instructions 

in the sequence provided. Replacement of parts should 

be required only occasionally and should not occur under 

normal pipettor use.

Problem

Cause

Solution

Droplets of 

liquid remain 

in the pipet 

tip.

The pipet tip is emptied 

too fast.

Decrease the speed 

of pressing the 

pipettor pushbutton.

The pipet tip wettability 

has increased due to 

extensive use.

Replace the tip with 

a new one.

Содержание Lambda EliteTouch

Страница 1: ...Corning Lambda EliteTouch Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...6051 0 5 10 20 6052 2 20 50 6053 5 50 100 6054 10 100 200 6055 20 200 1000 6056 100 1000 8 channel and 12 channel Pipettors 8 10 12 10 6057 6061 0 5 10 8 50 12 50 6058 6062 5 50 8 200 12 200 6059 606...

Страница 3: ...ector bar in the multi channel pipettors Materials Used Handle parts Polypropylene PP Polycarbonate PC Polyacetal POM Stainless steel Shaft and piston area parts Polypropylene PP Polyvinylidene fluori...

Страница 4: ...e the pipettor as it may damage the pipettor The single use of the tip reduces the risk of contamination of samples Keep the pipettor clean avoiding the use of aggressive substances e g acetone for cl...

Страница 5: ...ower position A B C The accuracy and precision are obtained gravimetrically using Corning pipet tips performing at least 10 measurements of distilled water at a temperature of 20 C 1 C according to th...

Страница 6: ...unter operating mode switch the calibration switch located under the CAL sign on the counter overlay This switch should only be used when the pipettor is calibrated upper position marked as CAL on the...

Страница 7: ...film remains relatively constant in successive pipetting operations with the same tip this error can be eliminated by forming the film before transferring the first sample This is done by aspirating a...

Страница 8: ...holding the handle unscrew the remaining part of the shaft holder set from it Then proceed according to the pipettor models shown below See ISO 8655 for the full table of the value of the conversion f...

Страница 9: ...d in the case of pipettor 2 L also the Seal P If you need to clean other components Unscrew the Shaft Cap M Remove from the Plunger I and pipetting Spring N Remove the Shaft J from the Connecting Bush...

Страница 10: ...g Spring N from the shaft Remove the Shaft J from the Connecting Bush L If necessary remove the Seal P from the Plunger I Check the condition of components and clean as recommended Replace any damaged...

Страница 11: ...ny damaged parts Assemble the components in reverse order When inserting the shaft or the shaft holder set into the Connecting Bush make sure that they are inserted all the way to the end by gently ro...

Страница 12: ...ealing elements e g after repeated autoclaving procedures Sterilization Sterilization in an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under other c...

Страница 13: ...Shaft Kit with seal and O ring 2 SP9811 1 J O Shaft kit with O ring 10 SP9812 1 20 SP9813 1 J Shaft 50 100 SP9814 1 200 SP9815 1 1000 SP9816 1 L Connecting Bush All SP9810 1 H Tip ejector 2 10 SP9817...

Страница 14: ...natural forces or other causes not arising from defects in the original material workmanship This warranty does not cover the PTFE seal O ring or shaft Claims for transit damage should be filed with t...

Страница 15: ...EUROPE France t 0800 916 882 f 0800 918 636 Germany t 0800 101 1153 f 0800 101 2427 The Netherlands t 20 655 79 28 f 20 659 76 73 United Kingdom t 0800 376 8660 f 0800 279 1117 All Other European Coun...

Страница 16: ......

Страница 17: ...2019 Corning Incorporated All rights reserved 3 19 CLS AN 525 H EliteTouch IM Corning Incorporated Daniszewska 4 03 230 Warsaw www corning com lifesciences Made in Poland...

Отзывы: