REMOTE CONTROL BUTTONS

 

 

 

 

Button 

AED-mode button function 

Manual-mode and monitoring-mode button 
function 

VES

VT Salve

SVES

 

For switching between On and Off for MOTION 
ARTEFACT. This function is only available during 
analysis. 

Pressing this button once in ECG mode will bring 
up a recording of a singular ventricular 
extrasystole on the display. Pressing this button 
after having pressed a function button will bring 
up either a self-terminating ventricular 
tachycardia or a supraventricular extrasystole on 
the display. 

 

For switching between On and Off for fitting the 
training electrode to the mannequin.  This 
function is only available when the training 
electrodes are connected to the AED Trainer.  

This button indicates to the device that the 
monitoring electrodes have been attached and 
makes it start showing a heart rhythm. 

NSR

SBrady

STachy

 

 

Pressing this button will cause all shock 
decisions to be changed to NO SHOCK 
ADVISED. This function can only be used if 
either the messages PRESS FOR ANALYSIS or 
ANALYSIS IN PROGRESS, STEP BACK! are being 
displayed. 

Pressing this button will bring up a normal sinus 
rhythm, a sinus bradycardia with approx. 40 
beats per minute or a sinus tachycardia with 
approx. 180 beats per minute. 

AVB 1°

Wenckeb.

Mobitz

 

No function 

Pressing this button will make the device 
simulate a first-degree AV block, a Wenckebach 
type 2° AV block and a Mobitz type 2° AV block. 

VVI

Exit-Block

VAV

 

Pressing this button once will activate a service 
alert, and will cause the message CALL 
CUSTOMER SERVICE to be displayed. To cancel, 

In ECG mode, this button will bring up various 
pacemaker rhythms: A single-chamber 
pacemaker, an exit block and a dual chamber 

Содержание Trainer 1000 SE

Страница 1: ...SE Supplementary user manual ONLY FOR USE IN CONJUNCTION WITH THE TRAINER 1000 USER MANUAL V 1 0 GBR ...

Страница 2: ... and training device for realistic Advanced Life Support training on a defibrillator AED and providing treatment to patients on the basis of an ECG diagnosis Due to the similarity of its design and operating elements the Trainer 1000 is particularly suitable for training and instructing people in the use of the Lifepak 1000 series defibrillators The Trainer 1000 does not deliver energy through adh...

Страница 3: ...ersonnel that is not expressly and legally authorised to perform electrotherapy in the country of application are advised to consult the head of their department before use IMPORTANT INFORMATION ON PERFORMING ECGS WITH THE TRAINER 1000 The LIFEPACK 1000 SE s monitoring function is designed to monitor the heart s electrical activity through a stable limb lead II cable The lead and the amplitude set...

Страница 4: ...to the device that the monitoring electrodes have been attached and makes it start showing a heart rhythm NSR SBrady STachy Pressing this button will cause all shock decisions to be changed to NO SHOCK ADVISED This function can only be used if either the messages PRESS FOR ANALYSIS or ANALYSIS IN PROGRESS STEP BACK are being displayed Pressing this button will bring up a normal sinus rhythm a sinu...

Страница 5: ...d to turn off the AED Trainer Pressing this button will bring up a bigeminus a 3 1 AV block or atrail fibrillation on the display To change to ECG mode Pressing this button will make the Trainer 1000 only use the signals from the connected lead for monitoring Pressing this button will make the device switch to ECG mode No function Pressing this button will enable you to activate the heart rhythms ...

Страница 6: ...hocks on the basis of the ECG analysis derived through the electrodes Instructors can chose the ECG recording to be shown on the display to provide advanced training in the use of additional and drug based treatment The heart rhythm can be changed at any time in manual mode If the Trainer 1000 is configured in such a way that the analysis function can be used in manual mode users will also be able...

Страница 7: ...ECG will now be shown On the remote control ECGs can be selected as follows Black ECGs Press button Red ECGs Press the red button first and then the ECG button Blue ECGs Press the blue button first and then the ECG button To exit ECG mode simply reverse these steps If the rhythm shown by the ECG is shockable the Trainer 1000 will prompt you to connect defibrillation electrodes and to exit ECG mode...

Страница 8: ...n shockable rhythm Based on non shockable rhythm Fix 1 SSSSK FFFFSr Fix1 Fix2 Fix2 KKKK AAAA Fix1 Fix2 Fix3 SSKKKK FFAASrSr Fix3 Fix5 Fix4 SSKSSK FFAFFSr Fix4 Fix5 Fix5 KSKSS AFSrFF Fix4 Fix5 Customer XXXXX S orr K F oder A Customer Fix4 Customer Fix5 S Shock advised K No Shock Advised Sr Sinus rhythm F Fibrillation A Asystole ...

Страница 9: ...able The designations of the ECG leads leading to the patient can vary but are generally as follows AHA designations IEC designations Electrode position RA R Right arm LA L Left arm LL F Left leg If your mannequin does not have any connections for the ECG cable you can use conventional adhesive ECG electrodes In this case we recommend using Ag AgCl solid gel adhesive electrodes and to leave them o...

Страница 10: ...ctrodes has been properly inserted into the socket on the AED Trainer and the instructor has verified that the electrodes have been correctly attached via the remote control If while using the AED Trainer the training electrodes are removed from the device it will display the message CONNECT ELECTRODES the training electrodes on the other hand come off the mannequin it will not display this messag...

Отзывы: