COM Server Commands
190
SCPI.SENSe(
Ch
).CORRection.COLLect.ACQuire.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel
Ch
,
Ch
Description
Measures the calibration data of the open standard for the specified
port.
Range
Port number is 1 or 2.
Out of Range
An error occurs.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN=
1
Notes
The property writing starts the measurement for the channel
independently of the trigger initiation and trigger source settings. The
function of the property writing waits for the completion of the
measurement.
Equivalent
Softkeys
Calibration > Calibrate > Response (Open) > Open
Calibration > Calibrate > Full 1–Port Cal > Open
Calibration > Calibrate > One Path 2–Port Cal > Open