![Copper Mountain Technologies PLANAR TR1300/1 Скачать руководство пользователя страница 181](http://html1.mh-extra.com/html/copper-mountain-technologies/planar-tr1300-1/planar-tr1300-1_programming-manual_2666189181.webp)
COM Server Commands
181
SCPI.SENSe(
Ch
).BANDwidth.RESolution
Object Type
Property (read/write)
Data Type
Double
Target
Channel
Ch
,
Ch
Description
The IF bandwidth value.
Range
from 10 to 30000
Resolution
In steps of 3. (10, 30, 100, 300, 1000, 3000, 10000, 30000)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10000
Unit
Hz (Hertz)
Syntax
Dim Value As Double
Value = app.SCPI.SENSe(Ch).BANDwidth.RESolution
app.SCPI.SENSe(Ch).BANDwidth.RESolution = 100
Equivalent
Softkeys
Response > IF Bandwidth