Maintenance
September 10, 2020 | 296742-ENG R00 17
4.3.1 Cleaning
Regular and thorough cleaning will help to ensure best performance.
•
Clean the chamber and machine compartment drain pans and drain lines regularly.
•
Clean the lamps regularly for maximum lighting efficiency.
•
Clean the internal walls regularly for maximum reflectance.
•
On air-cooled, self-contained chambers inspect to ensure the air-cooled condenser fins
are free of dust and debris, use a fine brush and vacuum to clean the fins or use
nitrogen or compressed air to clean the fins.
•
Avoid using high-pressure washers or high-pressure gases as this could cause damage
to the aluminum fins. Use maximum pressure 60 psi (4.14 bar).
Electrical power remains at the main terminals. Use extreme caution during
maintenance procedures to prevent injury.
Controller may come equipped with an optional Uninterrupted Power Supply (UPS)
such that power will remain live for a period of time even if the main power supply is
disrupted or turned OFF.
Use extreme caution to prevent injury. If your unit comes equipped with a UPS, it will
be located inside the control panel.
DO NOT service the control panel without using proper ESD procedures, including the
use of a grounding strap and/or anti-static mat.
Use only original replacement parts when maintaining and servicing the equipment.