Instrument Calibration
28
515 HC04 - 8 June 2007
Calibration Set Mode
In Calibration Set mode, you can change the settings of the “programmable”
parameters. You must enter the system password to change the setting of the
“password-protected” parameters.
Use the following procedure to enter Calibration Set mode:
1.
Press
to scroll to the
CAL MENU
prompt.
2.
Hold the
key.
The instrument beeps once, illuminates the
Cal
indicator and shows
CAL
on the display panel.
3.
Press
to select any flashing menu heading except
END
.
4.
Hold
for two seconds.
The instrument requests a password.
5.
Press
or
to change the value of the current digit. To select the
next digit, press
.
6.
Press
to accept the password.
•
The instrument makes two beeps for a correct password entry and
enables you to change the “programmable” and “password-protected”
parameters.
•
The instrument makes one beep for an incorrect password entry and
enables you to change only the “programmable” parameters.
The instrument illuminates both the
Cal
and
Set
indicators.
7.
Edit the instrument parameters as required. The programmable values
are indicated by the flashing display.
•
To change a numerical value, press
to increase a value, or press
to decrease a value. Press a key momentarily to change the value
one number at a time. Hold a key to scroll through the numbers. To
proceed to next digit, press
.
•
To change an option setting, press
or
to scroll through the
options.
8.
Press
to accept the currently displayed value and proceed to the next
parameter. You can press
to return to the main calibration menu.
9.
To exit from Calibrate Set mode, press
to scroll through the main
calibration menu to
END
, then press
. Otherwise, from any menu,
you can press and hold
for two seconds.
The instrument makes two beeps and cancels the
Cal
and
Set
indicators.
DISPLAY
SET
Run
Set
Alarm
Cal
SET
SET
Run
Set
Alarm
Cal
SET
DISPLAY
SET
SET
Run
Set
Alarm
Cal
Содержание 515
Страница 4: ...iv 515 HC04 8 June 2007 ...
Страница 8: ...viii 515 HC04 8 June 2007 ...
Страница 16: ...Introduction 8 515 HC04 8 June 2007 ...
Страница 60: ...Instrument Calibration 52 515 HC04 8 June 2007 ...
Страница 80: ...Communications 72 515 HC04 8 June 2007 ...
Страница 82: ...74 515 HC04 8 June 2007 ...
Страница 88: ...80 515 HC04 8 June 2007 ...