Installation & Maintenance
12
515 BR03 - 15 August 2021
Terminal Wiring Insulation
Terminals 1-30 are all rated at <35Vdc and wiring with basic insulation is
adequate.
Use copper conductors only.
Relay Wiring
Relay terminals are 31-35 or 31-36 depending on the version of option card
fitted. When controlling circuits powered by mains, or voltages >35Vdc, it
is necessary to use UL/CSA approved cabling with supplementary
insulation and a current capacity suitable for the connected circuit.
(0.75mm², 6A current capacity is sufficient).
Use copper conductors only. The maximum current must be <5A as stated
in the Specifications.
Mains Power Wiring
Only use UL/CSA approved cabling with supplementary insulation and
copper conductors. A minimum cable cross section area of 0.75mm² (18
AWG) is required. However, this must be selected with respect to the
installed overcurrent protection device and in accordance with the local
relevant Electrical Code of Practice as dictated by the Authority Having
Jurisdiction (AHJ).
The terminal designations (L/N/E) are clearly indicated just above the mains
input connector on the rear panel of the instrument.