INSTRUCTION   MANUAL 

 

CAT  1050 

– 1400 

STANDARD    INCUBATORS 

 
                                         
 
 
 
 
 

        CONTHERM SCIENTIFIC LIMITED 
     P O BOX 30-605   LOWER HUTT 5040 
 
                TEL:     (0064-4)  568 8034 
                FAX:     (0064-4)  568 8095 

 

            EMAIL:  [email protected] 

              
 
 
 
 
 

Содержание Cat 1050

Страница 1: ...INSTRUCTION MANUAL CAT 1050 1400 STANDARD INCUBATORS CONTHERM SCIENTIFIC LIMITED P O BOX 30 605 LOWER HUTT 5040 TEL 0064 4 568 8034 FAX 0064 4 568 8095 EMAIL contherm xtra co nz...

Страница 2: ...perating instructions have been observed and the defect is due solely to faulty design material and workmanship That the defective part be returned freight paid to the nearest sales service office Uni...

Страница 3: ...Manual7 1000_ZP21_0213 FEB 2013 2 STATEMENT of CONFORMITY This CONTHERM cabinet conforms to the following standards Electrical Safety Complies with AS NZS3350 1 1994 EMC Complies with EN 61326 1 1997...

Страница 4: ...troduction and Specifications SECTION 3 Operating Instructions SECTION 4 Theory of Operation SECTION 5 Customer Troubleshooting and Maintenance SECTION 6 Parts Lists for Spares and Accessories IMPORTA...

Страница 5: ...res obtained at one site and that at another site for those sites which give the greatest difference d TEMPORAL VARIATION The maximum value of the temperature range obtained for the standard site with...

Страница 6: ...orrosive resistant epoxy powder coated exterior Safety Fitted with an independent user adjustable Hi Limit completely separate from normal controls Convection All units come with mechanical convection...

Страница 7: ...e rear of the cabinet for air circulation NB Ensure ALL of the distinctly coloured cable ties If fitted inside cabinet are REMOVED BEFORE operating the incubator NB These cabinets are NOT suitable for...

Страница 8: ...control until the TIMER runs out of time The cabinet WILL NOT OPERATE if the TIMER has been set to 0 00 8 To look at the current TIME press the TIME button for 2 seconds the amount of time left will...

Страница 9: ...e FOR YOUR SAFETY To ensure safe operation the three pin plug supplied must be inserted ONLY into a standard three pin power outlet that is effectively earthed through the normal building wiring Exten...

Страница 10: ...g opened otherwise check the cabinet to determine if the Internal fan is still operating and that the MECHANICAL Hi Limit control is not interfering with normal operation 3 Preset alarm The cabinet ha...

Страница 11: ...sing the UP DOWN buttons until it agrees with the Calibration Thermometer NB If when attempting to press BOTH buttons together the temperature SET POINT adjusts either up or down it means you are NOT...

Страница 12: ...w SPATIAL PERFORMANCE The cabinet should be set to operate at the Contherm specified calibration conditions Check Spatial Variation 37o C Refer manual for specification 8 points as per diagram 6R 4C 3...

Страница 13: ...re in degrees centigrade OR elapsed time in hours and minutes 4 LED Indicators Left LED indicates when the heating element is ON Right LED indicates when controller is in SET mode 5 Refer Caution Inst...

Страница 14: ...air rotor such that the discharged air passes over the active portion of the element The chamber temperature is measured by a RTD SENSOR the resistance of which increases with increasing air temperatu...

Страница 15: ...of the interior base Any spillage or breakage within the cabinet should be cleaned up immediately with the cabinet switched off and unplugged at the mains NB These cabinets are NOT suitable for operat...

Страница 16: ...DEAD Check wall socket and wall socket circuit breakers B LARGE TEMPERATURE VARIATION IN CABINET CONTROLLER HEATER LED PULSING Internal fan not operating Check for air movement inside cabinet Cabinet...

Страница 17: ...tion and size of SPACERS e Remove PC Board List FULLY all fault details carefully pack return to Agent for repair ALWAYS supply contact details and FULL physical return address The refitting of the co...

Страница 18: ...Manual7 1000_ZP21_0213 FEB 2013 17 ELECTRICAL GP LAYOUT ZP21 1050 1400 Drawing 550 59...

Страница 19: ...GP1064 Hinges Glass Door GP1006 Glass door Catch Fan Motor E G O GP1000 Elements Cat 1050 1100 300W ELEMEGO300 Cat 1150 1200 300W ELEMEGO300 Cat 1300 1400 600W ELEMEGO600 Temperature Sensor 1000 RTD G...

Страница 20: ...x 507mm W Cat 1150 571mm H x 587mm W Cat 1200 636mm H x 700mm W Cat 1300 462mm H x 700mm W 2off Cat 1400 614mm H x 700mm W 2off NOTE If replacement doors are required hole placement and dimensions var...

Страница 21: ...spatch to our customers Contherm calibration equipment is traceable to a YSI Model 60 gallium melting point temperature standard CAT No APPLIANCE No DATE TEST RESULTS CALIBRATION FACTORS Test Temperat...

Страница 22: ...Lead Other Cabinet Undamaged when Packed When Received No dents scratches or obvious damage Certificate of Conformance Completed Factory test sheet attached Ensure cabinet has been fully tested and a...

Страница 23: ...nual SECTION 3 set for 37 0 o C Temperature stable 37o C Allow cabinet to stabilise at temperature for 1 hour Check Alarm Operation Open door for 30mins after cabinet stabilised Check Alarm Cancellati...

Страница 24: ...st PQ Leave with Cabinet CONTHERM CAT No App No Date Standard Incubator Operation PQChecked Comments Check Temporal Variation 37o C Refer manual for specification cabinet empty center only see verifyi...

Страница 25: ...r cabinet arrive in good condition YES NO 2 Any difficulties experienced in setting up YES NO Comments 3 Is this your first CONTHERM purchase YES NO 4 Any previous problems of a specific nature with C...