EKG6012
28
Mode
[Auto 6×2+1]/ [Auto 6×2]/ [Auto
4×3+1]/ [Auto 4×3]/ [Auto 3×4+1]/
[Auto 3×4]/ [Auto 2×6+1]/ [Auto
2×6]/ [Rhythm 7]/ [Rhythm 6]/
[Rhythm 5]/ [Rhythm 4]/ [Rhythm
3] and [Rhythm 2]/ [Manual]
The system takes the selected option as
default print mode.
Lead
Gain
[Smart]/[Current]
The option selected will be used as printing
gain mode. "Smart" means the system will
adjust gain automatically to fit paper
height; "Current" means it will use screen
waveform gain as that of printing.
Auto
Strip
[3 sec]/[4 sec]/[5 sec]/[6 sec]/[8
sec]/[10 sec]/[15 sec]/[20 sec]/[25
sec]
The system takes the selected option as the
print time length of each strip.
Rhyth
m
Strip
[10
sec]/[15
sec]/[20
sec]/[25
sec]/[30 sec]
When “Print Mode” is set to “Rhythm 12”
“Rhythm 10” “Rhythm 8” or “Rhythm 6”,
the system takes the selected option as the
print time length of each waveform.
Avera
ge
QRS
[6×2+
Mark]/[6×2]/[4×3+
Mark]/[4×3]/[3×4+
Mark]/[3×4]/[None]
When “Print Mode” is set to “Auto” or
“Rhythm”, the system uses the selected
format to print the average QRS waveform.
Auto-
Diag
[All]/[Data]/[Conclusion]/[None]
The diagnosis contains data and conclusion
two parts, which can be chosen by user as
demand.
Period
ic
[OFF]/[per 1 min]/[per 2 min] /[per
5 min]/[per 10 min]/[per 20
min]/[per 30 min]/[per 60 min]
During the ECG acquisition process, the
system will automatically activate the
printing operation according to the selected
time interval. When the printing mode is
manual mode, the printing will output
“Auto 12×1” format, otherwise, it will
output according to the current setting
mode.
Note 1: The settings of auto strip, rhythm strip, average QRS, auto diagnosis and
periodic print are only optional in auto mode and rhythm mode.
Note 2: If print time length is less than 8s, the sampling and analysis time will be 8s; if
the print time length is equal to or greater than 8s, the sampling and analysis time keeps
the same with print time.
7.9 Lead placement
In the main interface, click
to view the schematic diagram of lead placement, the