Manual Electrophoresis Power Supply, rev. 32/2018
9
Manual programming
Manual programming permits to set voltage, current, power and time limits for a simple routine
electrophoresis run. Parameters can be changed temporarily without interrupting the run.
This menu is automatically shown in the EV0220 after start-up.
Verify if the instrument is switched off.
Connect the electrophoresis tank to the instrument.
Switch the instrument on.
Select MANUAL and press
✓
:
(not EV0220)
Enter the desired voltage limit
and press
✓
:
Enter the desired current limit
and press
✓
:
Enter the desired power limit
and press
✓
:
Enter the desired timer unit (h or kVh)
and press
✓
:
h
The timer will count in hours and
minutes.
kVh
The timer will count in kilovolthours.
Enter the desired timer value
and press
✓
:
h
The value format is
hh:mm
.
kVh
The format has a value with 2 decimal
places..
Set to 00:00 when required to
continue until
/
is pressed or the
device is switched off.
Press
/
to start the experiment.
•
At any time, you can return to the main menu by holding
✓
pressed.
•
At any time, you can avoid programming unnecessary parameters. To do so, simply press
/
instead of
✓
after entering the last relevant value and the instrument will immediately start the
experiment keeping the other parameters to their previous values.
•
To ensure that the desired parameter is kept constant, we suggest to preset all other parameters to
its maximum.
•
The behaviour of the power supply during a procedure is described in the chapter
PROCEDURE?
MANUAL
METHOD
SETUP
MANUAL
VOLTAGE
300
V
MANUAL
CURRENT
500
mA
MANUAL
POWER
150
W
MANUAL
TIMER UNIT
h
MANUAL
TIMER
00:30
h