background image

Service Manual

English Version

Smoke Evacuation System

60-6852-ENG (-).qxd  2/1/99  12:40 PM  Page 1

Содержание 1000 SES

Страница 1: ...Service Manual English Version Smoke Evacuation System...

Страница 2: ...han intended 1999 CONMED Corporation Mailing Address 310 Broad Street Utica New York 13501 U S A Shipping Address 7211 S Eagle St Englewood Colorado 80112 U S A For Technical Services or Return Author...

Страница 3: ...spection 3 1 3 4 Troubleshooting the System 3 2 4 0 CUSTOMER SERVICE 4 1 4 1 Equipment Return 4 1 4 2 Ordering Information 4 1 4 3 Available Accessories 4 1 5 0 THEORY OF OPERATION 5 1 5 1 Filtration...

Страница 4: ...1 Control Panel 2 1 2 Back View 2 4 3 Front View 2 4 4 Air Flow Diagram 5 2 5 System Block Diagram 5 3 6 Exploded View 6 5 7 Insulation Assembly 6 6 8 Electrical Schematic 6 7 Figure Title Page...

Страница 5: ...subjected to prolonged expo sure The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides opti mal odor removal The fourth stag...

Страница 6: ...CHNICAL PERSONNEL There are no user serviceable components in the CONMED 1000SESTM Smoke Evacuation System Refer service to qualified service per sonnel Use only with the AC cord provided and always p...

Страница 7: ...WN arrow button decreas es suction The suction control should be set at the lowest setting to completely remove the surgi cal smoke from the operative site Each time either arrow button is depressed t...

Страница 8: ...le signs of structural damage 2 Insert the CONMED 1000SESTM Filter into the round opening in the front of the smoke evac uation system Be sure that the filter is seated completely against the rear of...

Страница 9: ...f surgical smoke When two indicators are illuminated on the suction con trol scale the suction control setting is the value between the two illuminated percentages 8 Deactivate the unit by a Pressing...

Страница 10: ...Breakers Remote Switch Jack Cautions Product Nameplate I 0 IPX1 Control Panel Filter Inlet Filter CONMED 1000SES Handle Cooling Vents Power Switch Remote Switch Jack Circuit Breakers Cautions Power Co...

Страница 11: ...ng it with a cloth that has been dampened not dripping with a disinfectant or a mild soap or detergent solution such as Windex or Formula 409 Windex is a registered trademark of the Drackett Products...

Страница 12: ...Tubing 4 Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual 1 Check power outlet and connection to rear panel of machine 2 Reset circuit breaker on bac...

Страница 13: ...001 Remote Switch Adapter 10 Cord Cat No 60 6886 002 Remote Switch Adhesive Pad Bag of 5 Cat No 60 6887 001 CONMED Disposable Smoke Evacuation Handpiece Case of 5 Cat No 60 6800 001 Hand Held Disposa...

Страница 14: ...4 2 This page intentionally left blank...

Страница 15: ...hose healthcare workers who breathe the gases For this reason the highest quality virgin coal grade activated carbon is used This activated car bon preferentially removes toxic organic gases rather th...

Страница 16: ...lled by a triac on the con troller PWB and runs on mains power This motor blower is rated at 115 CFM and a static suction of 97 H2O Replacement brushes for this AC motor are available from CONMED Tech...

Страница 17: ...5 3 Figure 5 System Block Diagram...

Страница 18: ...5 4 This page intentionally left blank...

Страница 19: ...mine the entire system for physical damage Contact BioMed or CONMED for service if damaged Yearly Inspections Perform the chassis ground integrity test per Section 6 3 Check the Service light on the m...

Страница 20: ...uum hose clogged 3 Check or replace hose 4 Motor Blower obstructed 4 Contact CONMED Technical Services 2 CONMED 1000 SES 1 Not plugged into electrical 1 Check power outlet or check to does not functio...

Страница 21: ...If necessary carefully remove the insulation from the base If any insulation is removed it may not be reinstalled because the adhesive may not func tion properly Newer models do not have adhe sive and...

Страница 22: ...luminate after 500 hours of use This is an indication to service the motor brushes Upon inspection if you find the carbon part of the motor brush is long or less replace the brushes Replacement brushe...

Страница 23: ...6 5 Figure 6 Exploded View...

Страница 24: ...6 6 Figure 7 Insulation Assembly...

Страница 25: ...6 7 Figure 8 Electrical Schematic...

Страница 26: ...h Jack Wire Harness 1 9 CHP C1X0 8 Microprocessor IC REV 1 1 10 WIRE ASSBLY Wire Assembly w E Terms 1 11 GASKET LED LED Gasket 2 12 MEMBRANEBLNK Membrane Switch Panel Blank 1 13 MEMBRANE CON CONMED Me...

Отзывы: