OPERATOR’S 

MANUAL 

Ver3.0C201-H

 

 

Fingertip Pulse Oximeter 

General Description 

Thank you very much for purchasing a Concord Pulse Oximeter. 

This  Manual  describes  the  Pulse  Oximeter’s  features  and  requirements,  functions,  specifications,  correct  methods  for 

transportation, installation, usage, operation, repair, maintenance and storage, as well as the safety procedures to protect both the 

user and equipment. Please read and follow the User Manual carefully before using this product. Failure to follow the User Manual 

may  cause  measuring  abnormality,  equipment  damage  and  human  injury.  The  manufacturer  is  NOT  responsible  for  the  safety, 

reliability and performance issues and any monitoring abnormality, human injury and equipment damage due to improper use. The 

specific products you received may not be exactly as described in this User Manual. This product can be used repeatedly for sports 

and aviation use. If you have any questions regarding to the use of this product, please call us at 888-970-2999 Monday- Friday

 

from 8:00 AM to 5:00 PM Central Time. 

Measurement Principle 

Principle  of  the  oximeter  is  as  follows:  A  mathematical  formula  is  established  making  use  of  Lambert  Beer  Law  according  to 

Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO

2

) in red and near-infrared zones. 

Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity

 

Pulse  Scanning  and  Recording  Technology,  so  that  two  beams  of  different  wavelength  of  lights  (660nm  red  and  940nm  near 

infrared  light)  can  be  focused  onto  a  human  nail  tip  through  a  clamping  finger-type  sensor.  A  measured  signal  obtained  by  a 

photosensitive element, will be shown on the oximeter’s display through process in electronic circuits and microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 

2. Red and Infrared-ray Receipt Tube 

Precautions For Use 

Before use, carefully read the manual. 

Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect 

user's safety and monitoring performance. It is recommended that the device should be inspected once a week. 

The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO2 measurement. 

Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and moisture. 

Do not use the fingertip pulse oximeter in an explosive atmosphere. 

Not intended for Continuous monitoring 

The fingertip pulse oximeter is intended only for sports and aviation monitoring 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 

intended for sterilization. 

Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, 

including batteries. 

10 

This  equipment  complies  with  IEC  60601-1-2:2007  for  electromagnetic  compatibility  for  electrical  equipment  and/  or 

systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical 

noise the environment, it is possible that high levels of such interference due to close proximity or strength of a source might 

disrupt the performance of this device. 

11 

Portable and mobile RF communications equipment can affect electrical equipment. 

12 

This equipment should not be used adjacent to or stacked with other equipment. 

13 

It may be unsafe to

 

—use accessories

detachable parts and materials not described in the instructions for use 

—interconnect this equipment with other equipment not described in the instructions for use 

—disassemble, repair or modify the equipment. 

14 

These materials that contact with the users skin contain medical silicone and ABS plastic enclosure all pass the ISO10993-5 

Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity. 

Contraindication 

It is not for continuous monitoring.   

Inaccurate measurements may be caused by 

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin). 

High ambient light. Shield the sensor area if necessary. 

Excessive user movement. 

High-frequency interference. 

Poor blood perfusion. 

Placement of a sensor on an extremity with a blood pressure cuff. 

Fingernail polish or false fingernails. 

Weak pulse quality. 

Low hemoglobin. 

Product Properties 

Simple to operate and convenient to carry. 

Small volume, light weight and low power consumption. 

High brightness OLED display SpO

2

, PR, and Pulse bar; two directions display mode. 

2 display modes. 

2pcs AAA-size alkaline batteries; battery-low indicator. 

When no or low signal is detected, the pulse oximeter will power off automatically in 8 seconds. 

Intended Use 

Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglob

in 

(SpO2) and pulse rate of adult and pediatric users in sports and aviation. 

Operation Instructions 

Install two AAA batteries according to the Battery Installation instructions. 

Place one of your fingers into the rubber opening of the pulse oximeter. 

Press the switch button one time on front panel to turn the pulse oximeter on. 

Keep your hands still for the reading. Do not shake your finger during the test. It is recommended 

that you do not move your body while taking a reading. 

Read the data from the display screen. 

Press the button again to toggle between two display modes. 

After turning on the Oximeter, each time you press the power switch, the Oximeter will switch to another display mode. 

There are 2 display modes shown as follows: 

1. 

                    2. 

 

Holding the power switch for longer than one second, will adjust the brightness of the oximeter. There are 10 levels of brightness. 

The default level is level four. 

 

 

 

 

 

Front Panel 

 

The pulse bar less than 30% indicates signal inadequacy and the displayed SpO

2

 and pulse rate value is potentially incorrect. 

Battery Installation 

1. 

Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) signs in the compartment. If 

the polarities are not matched, damage may be caused to the oximeter. 

2. 

Slide the battery door cover horizontally along the arrow shown as the picture. 

Note:   

 

Please remove the batteries if the pulse oximeter will not be used for long periods of time. 

 

Please replace the battery when the low battery power indicator. 

 

 

Using the Lanyard 

1. 

Thread thinner end of the lanyard through the loop. 

2. 

Thread thicker end of the lanyard through the threaded end before pulling it tightly. 

 

Warnings! 

 

Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards. 

 

Do not hang the lanyard from the device’s electrical wire. 

 

Please notice that the lanyard which is tied to the oximeter may cause strangulation due to excessive length. 

Maintenance and Storage 

1. 

Replace the batteries in a timely manner when low voltage lamp is lighted. 

2. 

Clean surface of the fingertip oximeter before it is used in diagnosis for users. 

3. 

Remove the batteries if the oximeter is not operated for a long time. 

4. 

It is best to store the product in -20

℃~+

55

 

and 

93% humidity

.

 

5. 

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage. 

6. 

Dispose of battery properly; follow any applicable local battery disposal laws. 

Cleaning the fingertip pulse oximeter 

Please  use  medical  alcohol  to  clean  the  silicone  touching  the  finger  inside  of  oximeter  with  a  soft  cloth  dampened  with 

70% isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test. 
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to 

dry thoroughly before reuse. 

The  use  life  of  the  device  is  five  years  when  it  is  used  for  15  measurements  every  day  and  10  minutes  per  one 

measurement. Stop using and contact local service center if one of the following cases occurs: 

 

An error in the 

Possible Problems and solutions 

is displayed on screen. 

 

The oximeter cannot be powered on in any case and not the reasons of battery. 

 

There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or 

the key is unresponsive or unavailable. 

Specifications 

1. Display Type 

OLED display 

2. SpO

2

 

Display range: 0%~100% 
Measurement range: 70%~100%   
Accuracy: 70%~100%: ±2%; 0%~69% no definition 
Resolution: 1%   

ARMS Value Analysis 

Testing Range

Data Points 

Bias 

ARMS 

Specification

70--<80 

81 

-0.70

1.82 

±2 

80--<90 

71 

-0.60

1.52 

±2 

90--100 

84 

-0.15

1.78 

±2 

70--100 

236 

-0.47

1.72 

±2 

Bland-Altman graphical plot of all sampled data points as below 

 

Note: A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to 

establish the SpO2 accuracy. The measured arterial hemoglobin saturation value (SpO2) of the sensors is compared to arterial 

hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in 

comparison to the CO-oximeter samples measured over the  SpO2 range  of 70%~100%. Accuracy  data is  calculated using the 

root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment

Particular requirements for the 

basic safety and essential performance of pulse oximeter. 

A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the specified calibration curve 

and the PR accuracy. 
The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3. 

3. Pulse Rate 

Display range: 0bpm~250bpm   
Measure range: 30bpm~250bpm 
Accuracy: 30bpm~99bpm, ±2bpm; 100bpm~250bpm, ±2% 
Resolution: 1bpm 

4. LED Wavelengths 

 

Wavelength 

Radiant Power 

RED 

660

±

3nm

 

3.2mw 

Concord Black Ox 

CCI-BLACK-OX