CONNECTING TO A REMOTE ALARM
There are two 8-pin remote alarm interface connectors located at the bottom of the Medical
IntelliSwitch™ cabinet marked ALARM A and ALARM B (See Figure 20). Having two Alarm
output connectors allows for connecting to two different external alarms which can be either
“Master” alarms or a CONCOA alarm. For convenience, the Medical IntelliSwitch is shipped with
two (2) cable connectors that mate to the alarm connectors on the cabinet. This allows for building
custom cable assemblies if needed. The recommended cable for these assemblies is 24 AWG, 8-
conductor stranded wire (Alpha # 1218C or equivalent).
To make a cable assembly cut the 8-conductor cable to length. Remove the outer jacket of the cable
to expose approximately 3/4 inch of the internal conductors on both sides of the cable. Strip away
1/4-inch of the insulation on each of the conductors.
On the 8-pin connector, slide the protective cover, cable retainer, and locking nut over the end of the
cable as shown in Figure 10.
Figure 10
Figure 11