Table 4 - Delivery Pressure Tolerance
Pressure
Tolerance
SW6-1
SW6-2
SW6-3
SW6-4
1
ON
OFF
OFF
OFF
2
OFF
ON
OFF
OFF
3
ON
ON
OFF
OFF
4
OFF
OFF
ON
OFF
5
ON
OFF
ON
OFF
6
OFF
ON
ON
OFF
7
ON
ON
ON
OFF
default
8
OFF
OFF
OFF
ON
9
ON
OFF
OFF
ON
10
OFF
ON
OFF
ON
11
ON
ON
OFF
ON
12
OFF
OFF
ON
ON
13
ON
OFF
ON
ON
14
OFF
ON
ON
ON
15
ON
ON
ON
ON
SETTING UNITS OF MEASURE
The Medical IntelliSwitch™ can display pressure in three different units of measure, PSI, BAR, and
MPA. The factory default is PSI. To change units of measure perform the following steps:
Note: In order for the DIP switches to be recognized, DIP switch SW4-3 must be OFF.
1.
Turn AC power OFF to the system.
2.
Open door by loosening the four screws holding it to the cabinet.
3.
Using Figure 8, locate DIP switch SW2 at the bottom of the electronic circuit board on the
door.
4.
Using Table 5, set the appropriate units of measure.
5.
Close system door and tighten 4 screws holding door to cabinet.
Figure 8