15
EMC Guidance
The
DQAPT
needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided.
This device has been thoroughly tested and inspected to assure proper
performance and operation.
Do not use the
DQAPT
simultaneously with other therapeutic
device (such as microwave), to avoid mis-operation.
Do not place or use the
DQAPT
nearby radio, television, copy
machine and fax machine.
Keep the
DQAPT
away from active HF surgical equipment and
the RF shielded room of a medical device for magnetic resonance
imaging, where the intensity of EM disturbances is high.
Use of the
DQAPT
adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If
such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) other than the
specified heart rate sensor to be used with the
DQAPT
should be
used no closer than 30 cm (12 inches) to any part of the
DQAPT
,
including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of the
DQAPT
could
result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation.