44
sputter or drip until all the air is removed.
•
Insulin will flow as a stream only when the
InPen is properly primed.
•
If the Injection Button is hard to push, the
needle may be clogged. Attach a new needle.
Repeat the priming steps until a stream of
insulin is seen.
If you are still unable to see a stream of insulin out of
the needle, do NOT use the InPen™. Contact Customer
Support for assistance or to obtain a replacement.
4. What should I do if I have an air bubble in the
Cartridge?
Priming your InPen™ will remove air. Hold the InPen™ so
that the needle is pointing up, and tap the Cartridge
gently with your finger so any air bubbles can collect
near the top. Repeat the priming steps until insulin is
seen. A small air bubble may remain in the Cartridge
after completion of the priming steps. If you have
properly primed the InPen™, this small air bubble will
not affect your insulin dose.
5. Why does no drop of insulin appear at the needle
tip when I check the insulin flow?
The needle may be blocked.
•
Screw on a new needle
•
Check the insulin flow until insulin appears at
the needle tip
13. Selecting Your Dose
1. Select your dose by turning the Dose Knob until the
desired dose is lined up with the Dose Indicator. Half
units are shown as lines between the numbers.
Содержание InPen
Страница 1: ...1 InPen System Instructions for Use...