16
IE-LGR-Uxxxx-16
U0111, U0121, U0141, U0141T
One-channel, two-channel
and four-channel thermometer
for external probes
These devices measure one, two or up to four
temperatures from external probes of the Pt1000 series.
The response to a temperature change depends on the
connected probe design. As a rule, this model is few
times faster than that using the internal sensor. It is often
used to monitor locations where the device itself is not
directly in the measuring area and only probes are in. The
maximum length of each probe lead should not exceed
15 m. It is recommended to use shielded cables. To
maintain the IP rating, the unused probe connectors
should be provided with a supplied closing cap.
The
battery in the device cannot be charged.
The models
U0111, U0121, U0141
have a high degree
of IP67 protection and probes with the connector
(Pt1000/E) can be connected to it. In order to maintain
the degree of protection, it is necessary to provide
unused connectors with the supplied cover.
The model U0141T
has a low degree of protection
IP20. The connectorless probes Pt1000/0 series are
intended to this model. The device has not a protection
against the ingress of water or dust. Do not use device
at places where such conditions can be expected.
The probes are connected to the pluggable terminal
block.
Connection:
The procedure for connecting the wires to the terminals
can be found in
U0111, U0121, U0141
U0141T
Содержание U0110
Страница 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Страница 18: ...18 IE LGR Uxxxx 16 Connection The procedure for connecting the wires to the terminals can be found in Annex 4...
Страница 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Страница 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Страница 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Страница 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Страница 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Страница 69: ...IE LGR Uxxxx 16 69 U8410...