IE-WFS-KAL-WiFi-01

 

 

 
 

Calibration (accuracy verification) 

To  verify  the  measurement  accuracy  (calibration)  of  temperature,  relative 
humidity, CO

2

 concentration  and barometric pressure  a specific equipment 

for generating these quantities (calibration chamber) is required. Calibration 
of  the  instrument  is  performed  by  placing  sensing  element  of  WiFi  sensor 
(DUT)  and  sensing  element  of  standard  into  calibration  chamber.  During 
calibration are values of DUT and values of standard compared. 

Calibration recommendations 

The calibration interval is determined by the user of the device. It is based 
on the manufacturer's recommendations and according to the requirements 
of  the  application  in  which  is  the  WiFi  sensor  used.  The  manufacturer's 
recommended calibration intervals are  stated in the User Manual. In some 
specific  applications,  calibration  must  be  performed  by  an  independent 
accredited laboratory. 

General calibration conditions 

  The  equipment  of  the  workplace  must  be  able  to  assess  the  measuring 

accuracy of the WiFi sensor by its parameters. As a rule, the instruments are 
checked for accuracy specified by the manufacturer, however the user of the 
WiFi sensor can determine his own parameters. For each measuring point it 
is necessary to calculate the so-called extended measurement uncertainty, 
which  includes  the  properties  of  the  whole  measuring  chain  (accuracy  of 
standards, inhomogeneity of the field in the calibration chamber, resolution 
of instruments, etc.). This uncertainty must be better than the required DUT 
accuracy in order to evaluate the measurement accuracy of the DUT. The 
determination of uncertainties is performed according to document EA-4/02. 
For  each  measuring  point,  the  standard  value,  the  measured  value  of  the 
DUT and the measurement uncertainty shall be reported. The evaluation can 
be added. 

  Used  standards  must  be  metrologically  traceable  to  the  valid  higher 

standards. A reference gas of known concentration (for CO

2

) or a calibration 

solution (for relative humidity) may to be used as the standard. There is an 
indicative  rule  which  says  that  standard  should  have  at  least  three  times 
better accuracy than DUT. 

  When  calibrating  physical  quantities, the  sensing  element  of  standard  and 

sensing element of DUT in calibration chamber should be placed closest as 
possible.  The  distribution  of  temperature  and  relative  humidity  in  the 
calibration chamber may not be completely homogeneous. Due to improper 
placement of sensing elements may be subject to a large systematic error. If 
the  calibration  is  carried  out  at different  temperature  than  ambient,  it  is 
necessary  to  ensure,  that  the  values  are  not  affected  by  the  heat  transfer 

Содержание W0710

Страница 1: ...www cometsystem com Calibration manual W0710 W0711 W0741 W3710 W3711 W3721 W3745 W4710 W5714 W7710 Manual for Calibration and Adjustment for WiFi sensors ...

Страница 2: ...ges in this manual without explicit agreement of company COMET SYSTEM s r o All rights reserved COMET SYSTEM s r o makes constant development and improvement of their products Manufacturer reserves the right to make technical changes to the device without previous notice Misprints reserved ...

Страница 3: ...CY VERIFICATION 5 Calibration recommendations 5 General calibration conditions 5 Preparing the WiFi sensor for calibration 7 Calibration procedure 7 ADJUSTMENT 8 WiFi sensor Adjustment Conditions 8 Adjustment Procedure 9 Saving of the new Adjustment 11 REVISION HISTORY 18 ...

Страница 4: ...e device Standard Measurement equipment e g thermometer or reference scale of certain quantity e g calibration gas used for calibration or adjustment Value Numerically quantified representation of measured physical quantity Sensing element Place of device or standard which is responsible to conversion of physical quantity to measured value Probe Sensing element at the cable or at detachable connec...

Страница 5: ...is necessary to calculate the so called extended measurement uncertainty which includes the properties of the whole measuring chain accuracy of standards inhomogeneity of the field in the calibration chamber resolution of instruments etc This uncertainty must be better than the required DUT accuracy in order to evaluate the measurement accuracy of the DUT The determination of uncertainties is perf...

Страница 6: ...ed the measuring range of the WiFi sensor The time for starting and stabilizing the calibration chamber must be known in advance or can be reliably ascertained during calibration This can be done by monitoring the values of the standard Stability of the calibration chamber means the condition when the calibrated quantity does not change anymore and the air circulating in the chamber has the same t...

Страница 7: ... counted from the moment when the controlled quantity has stabilized in the calibration chamber This time should be at least three times the response time t90 specified in the User manual for the calibrated quantity After this time read the measured value of the standard and the value from the calibrated DUT Set the chamber to generate another value for next point and repeat the above procedure fo...

Страница 8: ...ans of the WifiSensorUtility software WiFi sensor Adjustment Conditions Any unqualified entry into the instrument adjustment mode to may rewrite the existing calibration constants It can cause that the instrument will be unable to measure Follow recommendation and conditions for calibration during the device adjustment Adjustment of the measured values should be done in two point These points shou...

Страница 9: ...nded to back up the current sensor settings to file for possible restoring Save the backup file Backup file allows you to return the settings to the previous state including calibration Restoring can be done using Service Restore configuration from file menu item Note Calibration of Digi probes cannot be restored in this way see below Be aware that administrator password for device is required in ...

Страница 10: ...ity pressure and CO2 is also placed in the calibration chamber During adjustment proceed in the same way as described above for calibration Close and run the calibration chamber Set to calibration chamber to a first adjustment point value Wait for the device to stabilisation The device stabilization time can be counted from the moment when the controlled quantity has stabilized in the calibration ...

Страница 11: ...0 RTD sensors are very stable From this reason is not recommended to change calibration constants if the probe does not measure accurately There are two ways to improve measurement accuracy It can be set correction to cable length or user calibration can be used At first step enter Device settings and show Channels bookmark In case of probes with longer cables are used it should be set correction ...

Страница 12: ...12 IE WFS KAL WiFi 01 If correction to cable length is correct it can be set user calibration If the WiFi sensor measured 25 C before this setting then 26 C will measure after it ...

Страница 13: ... downloaded from device Digi probes can be restored into factory defaults only select Service Digi probe adjustment menu item select Input and click Read from probe button Temperature adjustment Is usually not necessary to adjust the temperature of Digi probe For adjustment enter the measured values of the standard and the device Click Save new calibration button finally If the device measured 25 ...

Страница 14: ...e new calibration button finally If the device measured 73 2 RH before adjustment then measures 75 3 RH after it Important note Capacitive humidity sensors usually drifts at high humidity and not at low humidity values In most cases it is sufficient to perform a one point adjustment in higher humidity as show at image ...

Страница 15: ...ct Service Device adjustment menu item two point calibration select Barometric Pressure channel 1 Click Enter Adjustment 2 Enter standard value for calibration point No 1 3 Enter device value for calibration point No 1 4 Enter standard value for calibration point No 2 5 ...

Страница 16: ...lue for calibration point No 2 6 Save new calibration to device 7 CO2 concentration Adjustment W4710 W5714 select Service Device adjustment menu item one point offset can be used select CO2 concentration channel 1 Click Enter Adjustment 2 ...

Страница 17: ...IE WFS KAL WiFi 01 17 Enter a zero value 3 Enter a zero value 4 Enter standard value for calibration point 5 Enter device value for calibration point 6 Save new calibration to device 7 ...

Страница 18: ...18 IE WFS KAL WiFi 01 Revision history Document version Date Note IE WFS KAL WiFi 01 January 2021 Initial document version Note Page numbers may to differ between document versions ...

Отзывы: