![Comet System T0610 Скачать руководство пользователя страница 45](http://html.mh-extra.com/html/comet-system/t0610/t0610_user-manual_2640750045.webp)
IE-SNC-Tx6xx-07
45
Operating humidity and atmospheric pressure range:
Type
Humidity range
Atmospheric pressure range
T0610
0 to 100%RH
without condensation
600 to 1100hPa
T4611
T3610
T3611
T7610
T7611
T7613D
T6641
850 to 1100hPa
T5641
T6640
0 to 95%RH
without condensation
T5640
Protection:
Type
Case with electronic Sensor cover
(or external probe with cable)
CO
2
probe
T0610
IP30
IP40
-
T3610
T7610
T6640
T7613D
T3611
T7611
T6641
IP65
T5641
-
T5640
-
T4611
depends on used Pt1000 probe
Other environmental conditions (external influences) according to HD 60364-5-51:
normal
Storage conditions:
same as the operating ranges
Working position T3611, T4611, T7611, T5641, T6641:
arbitrary
Working position T0610, T3610, T7610, T7613D, T6640, T5640:
with sensor cover downwards
End of operation
Disconnect the device and dispose it according to current legislation for dealing
with electronic equipment (WEEE directive). Electronic devices must not be
disposed of with your household waste and need to be professionally disposed.