IE-SNC-Tx6xx-07
28
Address [DEC] Address [HEX] Value
Type
20504
0x5018
Max. relative humidity memory
Int*10
20505
0x5019
Min. computed value memory
Int*10
20506
0x501A
Max. computed value memory
Int*10
20507
0x501B
Min. atmospheric pressure, CO
2
concentration memory
Int*X
20508
0x501C
Max. atmospheric pressure, CO
2
concentration memory
Int*X
Type:
DEC
register is in range 0
–
4500 (16bit)
BCD
register is in BCD format (16bit)
HEX
number in HEX format (16bit)
String
two characters in one 16bit register:
no
–
no alarm
hi
–
value is lower than set limit
lo
–
value is higher than set limit
Int*10
register is in format integer*10
–
16 bits:
(125=12.5°C;
error = 9999 or -9999)
Int*X
register format depends on value type (error = -9999 or -9998):
hPa
–
integer*10 (9760 = 976.0hPa)
PSI
–
integer*1000 (14156 = 14.156PSI)
inHg
–
integer*100 (2882 = 28.82inHg)
mBar
–
integer*10 (9761 = 976.1mBar)
oz/in
2
–
integer*10 (2265 = 226.5oz/in
2
)
mmHg
–
integer*10 (7321 = 732.1mmHg)
inH
2
O
–
integer*10 (3919 = 391.9inH
2
O)
kPa
–
integer*100 (9761 = 97.61kPa)
CO
2
–
integer (1050 = 1050ppm)
SOAP
The device allows you to send currently measured values via
SOAP v1.1
protocol. The device
sends values in XML format to the web server. The advantage of this protocol is that
communication is initialized by the device side. Due to it is not necessary use port forwarding.
If the SOAP message cannot be delivered, warning message via SNMP Trap or Syslog
protocol is sent. There are two formats of SOAP message. One for devices with CO
2
concentration measurement (T6640, T6641, T5640, T5641) and another for other devices.
SOAP message from T6640, T6641, T5640, T5641
Structure of SOAP messages is described at XSD schema. It can be downloaded from:
http://cometsystem.cz/schemas/soapTx5xxCO2.xsd. SOAP message example: