CDX2000 / 13.56 MHz & 400 kHz Manual
© 2006 Comdel, Inc.
9
3
Unpacking and Inspection
1.
Carefully unpack the unit and inspect for any obvious signs of physical damage that might have occurred
during shipment.
2.
Check the outside of the unit for missing or loose mounting screws or broken parts.
3.
If there is shipping damage or the unit fails to operate properly upon receipt, report damage to the carrier
immediately and notify Comdel factory within 30 days of receipt of unit. Failing to report any damage within
this time period is the same as acknowledging that the unit was received undamaged.
4.
If returning the product for repair you must:
•
Return the unit in the original, or equivalent, shipping container
•
Receive a Return Materials Authorization (RMA) number from the factory prior to the return of the
product to Comdel for repair
•
Place RMA numbers clearly on the shipping container or on the packing slip
CAUTION: Breaking the seal or removing the warranty decal from this unit will void the warranty. If internal
damage is suspected, contact factory for assistance.
4
Maintenance
The CDX2000 is designed to operate unattended for long periods of time. Should periodic maintenance or service
be required, it should only be performed by qualified service persons. Contact the factory for maintenance or
service recommendations.
5
Preparation for Use
A. Line Requirements
The CDX2000 is designed to operate from a 208 VAC, three phase line. The system will still function within
specifications when the line voltage fluctuates between 187 volts and 229 volts. Voltages over the
recommended 208 VAC, however, reduce the safe performance margins designed into the system and should
be avoided. The system draws a maximum of 12 Amps per phase when used to drive a load of 50 ohms.
Under conditions of mismatch, the amplifier could draw slightly more current.
Nominal Input:
208 VAC, 3 Phase
50/60 Hz
20 Amps
B. Cooling
Requirements
The ambient air temperature should not exceed 40
°
C. There should be enough room over the top of the
amplifier and along the sides to permit an unobstructed airflow through the unit. Water temperature should not
exceed 25
°
C, and should not fall below a temperature where excess condensation could develop within the
unit. Water volume should be at least 6 l/m (1.6 GPM) and pressure kept below
414 kPa (60 PSI).