background image

 

Page 4 of 11

 

INSTRUCTION MANUAL FOR LI-3100

 

LINE IMPEDANCE STABILIZATION NETWORK (LISN) 

 

2.0 Product Description 

 

Overview 

 
The  LI-3100  Line  Impedance  Stabilization  Network  was  specifically  designed  for  providing 
standardized line impedance during Electromagnetic Compatibility (EMC) testing in accordance 
with 

RTCA DO-160, MIL-STD 461F & CISPR 25 

requirements. In addition, it minimizes the noise 

generated by equipment other than the EUT to be included in the measurement. 
 
The  LI-3100  is  shipped  with  two  LISN  housed  in  separate  enclosures  for  single  phase  power 
applications.  For wiring convenience, each LISN has either a black or red shrouded banana for 
both power and EUT connection. The following items are included with each LI-3100 LISNs. 
 

Standard items included: 

Equipment, accessories, and documents supplied with the LI-3100 LISN are as follows: 

o

 

Calibration data and Certificate traceable to NIST.

 

o

 

Superior Electric shrouded 100A plugs for EUT and power connection.

 

 

Optional items 

o

 

ISO-17025 calibration data and certificate. 

 

 

 
 

 

Distributed by: Reliant EMC LLC,  +1 408 916

5750,  [email protected], www.reliantemc.com

Содержание LI-3100

Страница 1: ...FOR LI 3100 LINE IMPEDANCE STABILIZATION NETWORK LISN INSTRUCTION MANUAL For LINE IMPEDANCE STABILIZATION NETWORK Model LI 3100 10 kHz to 400 MHz Distributed by Reliant EMC LLC 1 408 916 5750 info re...

Страница 2: ...ents 1 0 Introduction 3 2 0 Product Description 4 3 0 Product Specifications 5 4 0 Important Safety Precautions 6 5 0 Warranty 7 6 0 Front and rear panel marking 8 7 0 LISN Theory 9 8 0 Typical Perfor...

Страница 3: ...intenance and warranty information It also includes some basic guidance on LISN operation Information contained in this manual is the property of Com Power Corporation It is issued with the understand...

Страница 4: ...asurement The LI 3100 is shipped with two LISN housed in separate enclosures for single phase power applications For wiring convenience each LISN has either a black or red shrouded banana for both pow...

Страница 5: ...ing Forced air Power Adapter for fan Output 12 VDC 1 Amp Input 120 VAC 60 Hz 230 VAC 50 Hz Fuse 1 Amp 250 V T Adapter output connector Diameter 2 1 x 5 5 mm Center Positive Operating Temperature 5 C t...

Страница 6: ...n must not be negated by the use of a cheater plug or an extension cord without a protective conductor Caution The LI 3100 LISNs use forced air system to keep the air core inductors cool during the te...

Страница 7: ...either directly or indirectly from the use or handling of this product or any inability to use this product seperately or in combination with any other equipment When requesting warranty services it i...

Страница 8: ...e right of the EUT Power Port for connecting the Spectrum analyzer or EMI receiver for making conducted noise measurements 6 4 Grounding bolt GND The bolt marked GND are found on both front and back p...

Страница 9: ...imal effect This filter comprises of single stage low pass LC filter Therefore it has capacitor and an inductor for the filter The inductor used in the LI 3100 is air core type to eliminate the possib...

Страница 10: ...pared to DO 160 requirements Figure 3 Impedance AE Port Short compared to CISPR 16 1 2 CISPR 25 requirements 0 1 1 10 100 0 01 0 1 1 10 100 1000 Impedance Frequency MHz 0 10 20 30 40 50 60 70 0 1 1 10...