Sterilizers Bravo G4
- TECHNICAL MANUAL
108
TECHNICAL MAINTENANCE
EN
8. TECHNICAL MAINTENANCE
In addition to correct use, the user performs regular basic maintenance in order to guarantee efficient operation over
the device’s entire life.
In order to grant full device efficiency, additional routine checks with regular periodic check-ups must be performed
by Technical Service. It is also fundamental to perform a
periodic sterilizer validation
, i.e. a check of process
thermo-dynamic parameters and their comparison with the reference values detected by duly calibrated tools. Refer
to chapter 'GUIDELINES-SAFETY TEST-VALIDATION' included in this Technical Service Manual.
8.1. MAINTENANCE PROGRAMME
The table summarises the maintenance interventions required to maintain the sterilizer efficiency and performances.
Relevant Warnings messages, displayed by the unit, recall to some maintenance activities in charge to User and/or
Service.
SCHEDULE
ACTIVITY
PERFORMED BY
DAILY
Clean the Door Gasket and the internal part of the door
USER
WEEKLY
Clean external surfaces.
Clean the sterilization chamber and its accessories.
Clean the anti-dust filter.
USER
MONTHLY
Clean the internal water tank (ref. User Manual).
USER
PERIODICALLY
Warnings related to Scheduled Maintenance are
displayed with the following frequency.
WARNING MESSAGES
FREQUENCY
Chamber Filter Cleaning.
250 cycles or 3 months.
USER
Door Lock Lubrication.
250 cycles or 3 months.
USER
Anti-Dust Filter Cleaning.
500 cycles or 3 months.
USER
Bacteriological Filter
Replacement.
500 cycles or 6 months.
USER
Door Gasket Replacement.
1000 cycles* or 1 year.
SERVICE
General Service.
3000 cycles ** or 3
years.
SERVICE
YEARLY
Sterilizer validation (chapter 'GUIDELINES-SAFETY
TEST-VALIDATION') ***.
Door Gasket Replacement*.
Clean the internal water tank.
SERVICE
* Replacement every 1000 cycles or 1 year, whichever is first.
** Replacement every 3000 cycles or 3 years, whichever is first.
***Depending on local guidelines or regulations.
Содержание SciCan BRAVO G4
Страница 2: ......
Страница 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Страница 12: ...Sterilizers Bravo G4 TECHNICAL MANUAL 10 GUIDELINES SAFETY TEST VALIDATION IF REQUIRED EN 3 2 CONFORMITY REPORT...
Страница 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Страница 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Страница 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Страница 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Страница 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Страница 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Страница 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Страница 91: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN TECHNICAL SHEETS 89 1 Steam generator complete with heat insulation...
Страница 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Страница 116: ......