Fault
Description
Machine condition
Solutions
E170
Error door lock
•
Close door completely
•
Check the door lock
•
Check the connection of the door lock
•
Check door motor
E180
Door is not open
•
The door is not open
•
Open door with the DOOR button
•
The mixing container is clamped
•
Check the door lock
•
Check the connection of the door lock
•
Check door motor
E190
Fuse blown
•
Turn the machine off and wait one minute until the
fuse can be pushed in again.
7.2
E080 MU unlocked
The mixing unit was not positioned and fixed in the 12o'clock position.
This condition occurs upon actuation of the EMERGENCY STOP button or upon interruption of the
power supply during the mixing process.
AUTO
With the container
clamped
or the mixing unit
not loaded
, the mixing unit can be automatically
positioned and fixed with the help of
Button I (Auto)
MANU
With the mixing container unclamped
, the mixing unit must be positioned manually.
Upon actuation of Button III (MANU), the door can be opened by pressing the DOOR button.
Remove mixing unit from the mixing space and fix mixing unit manually in 12o'clock position
Quit manual positioning by pressing the E button
MU UNLOCKED
Press Button
for positioning
AUTO
MANU
MU UNLOCKED
Press DOOR
for opening Door
MU UNLOCKED
wait for lock MU
MU UNLOCKED
MU locked
Close Door
Press E to quit
MU UNLOCKED
Press Button
for positioning
AUTO
MANU