Collins EAGLET Скачать руководство пользователя страница 1

INSTRUCTION MANUAL FOR THE COLLINS EAGLET

Pulmonary Diagnostic Spirometer

#762001

Version: February 2003

© Copyright 2003 Collins Medical, Inc.

Caution

Federal law restricts this device to

sale by or on the order of a physician.

Содержание EAGLET

Страница 1: ...MANUAL FOR THE COLLINS EAGLET Pulmonary Diagnostic Spirometer 762001 Version February 2003 Copyright 2003 Collins Medical Inc Caution Federal law restricts this device to sale by or on the order of a...

Страница 2: ...btained for the return Rubber parts and other parts which are subject to rapid deterioration over which the Company has no control are warranted for ninety 90 days Collins Medical Inc shall not be lia...

Страница 3: ...ition 14 For Microsoft Windows 2000 16 Preparing the Eaglet for Calibration and Verification 19 Preparing the Eaglet for Testing 19 SECTION TWO PERIODIC MAINTENANCE Warning 21 Customer Support 21 Rout...

Страница 4: ...9 Advanced Power Management Settings for Notebook Computer Windows XP 16 1 10 Power Management Settings for Notebook Computer Windows 2000 17 1 11 Advanced Power Management Settings for Notebook Comp...

Страница 5: ...0 software manual which describes configurable aspects of the software patient testing report and database functions Each system also has its own hardware manual You are currently reading the Eaglet H...

Страница 6: ...Humidity 10 to 90 non condensing Auto Zero Yes Type of Pneumotach Silverman low thermal mass single screen pneumotach Linearity 2 Resistance 0 5 cm H2 O L sec Accuracy 3 Range 15 L sec Computer Hardw...

Страница 7: ...of IEC 60601 1 2 2001 09 2nd Ed If for some reason you are having problems with interference contact the Service Department at 800 635 3200 Examples of occurrences that could interfere with testing in...

Страница 8: ...at could result in damage to equipment Following are symbols that may be found on the Eagletand their meanings 2 X oC X oF Y o C Y oF High Voltage Attention Consult accompanying documents Protective E...

Страница 9: ...page 9 2001 05 30 Figure 1 2 Symbols that May be Found on the Eaglet Indicates the date by which the product must be used in the format Year Month Date Heavy Weight Fragile Keep Dry Transport and sto...

Страница 10: ...t of equipment within the department start this process immediately to avoid prolonged installation delays Unpacking 1 The Eaglet system comes packaged in 1 box which is identified by the serial numbe...

Страница 11: ...t so as not to be damaged Cable and Power Cord Connections 1 The following cabling instructions and accompanying diagram represent a sample cabling configuration Keep in mind that depending on the com...

Страница 12: ...ble into the adapter reserved for the monitor on the rear of the computer Secure the screws at both ends of this and all other connectors with a small screwdriver 8 Plug one end of the printer cable i...

Страница 13: ...l outlet This requires that each component be powered up individually by turning on the power switch The Eaglet receives its power from the USB port on the computer and will be powered up when the com...

Страница 14: ...must be disabled o The operating system s lid closing settings on the notebook computer must also be set to do nothing rather than go on standby or hibernation o The computer should never be put in st...

Страница 15: ...dialog box select Power Schemes 4 Set the system stand by options to Never 5 Set the system hibernates options to Never Figure 1 8 Power Management Settings for the Notebook Computer Windows XP 6 Clic...

Страница 16: ...by mode settings must be disabled o The operating system s lid closing settings on the notebook computer must also be set to none rather than go on standby or Power off o The computer should never be...

Страница 17: ...ns 3 In the Power options dialog box select Power Schemes 4 Set the Power Schemes options to Always On Figure 1 10 Power Management Settings for the Notebook Computer Windows 2000 5 Click the Advanced...

Страница 18: ...1 11 Advanced Power Management Settings for the Notebook Computer Windows 2000 7 Click the Hibernate Tab 8 Disable the hibernate Support option Figure 1 12 Disabling the Hibernate Support Button for t...

Страница 19: ...ill be held during testing Also the Eaglet should be stationary during calibration verification Figure 1 13 Calibration Verification Setup for the Eaglet Preparing the Eaglet for Testing 1 Insert the...

Страница 20: ...ient Holding the Eaglet during Testing 4 Connect the patient to the free end of the filter making sure he she maintains a tight seal with his her lips 5 For instructions on conducting spirometry tests...

Страница 21: ...epartment FAX 781 843 4024 or your local distributor Routine Cleaning of the Eaglet Handle Frequency As needed 1 Turn off the power to the computer and disconnect the Eaglet from the computer 2 All ex...

Страница 22: ...s infection control committee However when contamination is believed to have occurred such as with contagious patients it is acceptable to clean the pneumotach after each patient In addition the pneum...

Страница 23: ...to be soaked and swished around in solution but not scrubbed Scrubbing the screen may deform it and affect results Thoroughly rinse the pneumotach and allow it to dry as any residue or water clogging...

Страница 24: ...nly trained staff may use Cidex Wear gloves goggles and an apron or gown Avoid skin contact The solution is caustic and can cause skin irritation 2 Activate solution solution turns green when activate...

Страница 25: ...as holes in it or seems to sit loosely in the pneumotach core it must be replaced If the pneumotach cannot be calibrated or verified it must be replaced NOTES If the pneumotach needs replacing the Eag...

Страница 26: ...able PFT Kit includes DC 2 Filter ComfortFit Mouthpiece and Soft Noseclip box of 100 022242 Disposable Comfort fit Mouthpiece 02240x Disposable Cardboard Mouthpieces 021231 Disposable Soft Noseclip pa...

Страница 27: ...AC power connections for the monitor etc do not apply Problem 1 Nothing displays on the monitor screen or the PLUS program fails to boot up when the computer is turned on Check to make sure that 1 Th...

Страница 28: ...accompanying your monitor for further details on determining monitor problems Problem 3 The monitor displays a message such as Communication Error This is an indication that the computer and the Eagl...

Страница 29: ...ion i e you can t get it to Pass 1 In the Calibration section of the PLUS 2000 Manual refer to the part about pneumotach calibration Read about the error messages you will get if you do not conduct th...

Страница 30: ...a problem attach a flexible mouthpiece to the filter 5 Setup was correct prior to testing 6 The test was started properly according to the PLUS 2000 manual 7 The test was ended properly 8 The correct...

Страница 31: ...iately and unplug all AC power cords from their outlets Do not reapply power until all water or moisture is completely eliminated Contact Collins Medical Inc if necessary d Is the problem mechanical i...

Страница 32: ...source of the problem is found a Can it be fixed without outside help b Is enough information available to discuss the problem over the telephone with a service representative c Is an on site visit fr...

Страница 33: ...to type some commands at the keyboard or make minor adjustments relay detailed information during operations or try other troubleshooting steps at the system itself d For all third party software such...

Страница 34: ...Eaglet Manual Section Three page 34 Collins Medical Inc 220 Wood Road Braintree MA 02184 2403 781 843 0610 800 225 5157 FAX 781 843 4024 www collinsmedical com...