Interval
Test Name
Description
OFF
Alarm buzzer ON/OFF, CONT, TIME
ADJUST, INITIAL INFLATION
ADJUST
Alarm buzzer ON/OFF, Continued, Adjust Time,
Date and Initial Inflation Controls all Smart
Features if Applicable
CON
Safety Timer Test
Tests the “Watch Dog Timer Circuit”
1*
Telemeter Checker Mode
N/A
2*
TRM Test
N/A
2.5
Rear Panel SW Test, DIP Switch
Indicator, Front Panel Button Test
-Recommended for PM Verification-
Tests all rear panel switches (Adult/Neonatal,
Display, Printer), Front Panel Buttons (except
ON/OFF), Shows position of internal DIP
switches
3
Air Leakage Test
-Recommended for PM Verification-
Tests air leakage rate within the system
5
VER#, All Ram Clear
Verifies software version number. Clears all
settings stored in RAM (including the systolic
alarm settings).
10*
Telemeter Channel Test
N/A
15
Pump-Up Test
Automatically inflates the connected dummy cuff
and calculates the rate of inflation
20
RS232C Test
VerifiesTransmitt and Receive lines of the RS232
30
Printer Test
Commands the printer to print out a test pattern
45
Initial Deflation Test
-Recommended for PM Verification-
Automatically inflates and tests the deflation
parameters of the monitor (Counts, Maxpress)
60
Deflation Test
Automatically inflates the connected dummy cuff
can, the static deflation rate is measured and then
displayed
120
VER#, Display Test
-Recommended for PM Verification-
Verifies software version number. Tests all
elements of front panel display
180
Static Calibration Test
-Recommended for PM Verification-
Allows user to verify the accuracy of the internal
pressure sensor, and Overpressure settings
180__________
Software Overpressure Test
Systolic high alarm setting 240
Systolic low alarm setting 150
Allows user to verify the accuracy of the
software Overpressure setting.
7.2
Test Menu
* These tests can only be completed on Press-Mates with the telemetry option.
7-2