24
TECHNEWORKS PC SOFTWARE
The FSB-200-P-AC is specially designed to allow the operating temperature of the fluidized bath to be adjusted
from a remote source while the bath is unsupervised. Fitted to the rear of the fluidized bath is an RS232 port
which connects directly to the Eurotherm 2408 Comms1 RS232 module. The TechneWorks software is available
to download from the Techne website.
http://www.techne-calibration.com/adminimages/TechneWorksv2.zip
Software Features:
• Specify either °C or °F.
• Specify ramp rates.
• Log data from the instrument while connected to the computer and export the data to an Excel®
spreadsheet.
• Open, save, view and print logged data.
• Run a program in real-time mode.
• Specifying the logging interval from every 5 seconds to 60 seconds.
• Program daily start and stop times to automatically turn unit off and on for weekly schedules.
To set up remote operation:
1. Install the TechneWorks Software. We highly recommend that TechneWorks be installed on a PC that is
typically not restarted at night for updates or Antivirus scans, though the software does have an auto restart
feature if the PC is rebooted.
2. Connect the FSB-200-P-AC to an available coms port on your PC noting the port number.
3. Launch Techneworks and then select “Fluidized bath” as shown below.
4. The home screen will be displayed; click on “start device”.
5. From the “Serial Port” window select the Com port you are connecting to and click OK.
Содержание FSB-200 Series
Страница 1: ...Instruction Manual 7002757 CPB Version 12 1 Cole Parmer FSB 200 Series Fluidized Sandbath...
Страница 2: ......
Страница 34: ...NOTES 33...