52412Mechnicaloven_ver_1.2
48
4.3.6
Auto Run
When power is turned off and restored after more than 30 seconds, Auto Run function makes the
user set the function to determine whether or not to operate by remembering the state before the
power turns off.
Item
Function
Input Value
Default
Auto Run
When the power is turned off for more
than 30 seconds during temperature
control and restarts, set the function to
automatically run at the previously set
temperature.
Off / On
OFF
If the user sets the Auto Run and the power is restored after the device is turned off for 30
seconds or more during temperature control, follow the procedure below.
Previous
States
Operation in Basic Control
Auto Run Off
Auto Run On
Stopped
Stop
Stop
Running
Stop
Operation
Wait ON
Timer
Stop
The elapsed time of the Wait ON Timer is not be
restored and returns to the initial setting.
Wait OFF
Timer
Stop
The elapsed time of the Wait OFF Timer is not
restored, and starts from the beginning.
※
Refer to [4.2.5 Instantaneous Outage Recovery Function] for operation by instantaneous
power failure (recovery before 30 seconds).
※
Safety Lock Door function is not embedded in the OVF-800 Series.
Содержание 52412-15
Страница 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...
Страница 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...
Страница 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...
Страница 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...
Страница 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...
Страница 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...
Страница 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...
Страница 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...
Страница 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...