52412Mechnicaloven_ver_1.2
43
•
If the Wait ON Timer and Wait OFF Timer are both set, the activated timer and
the remaining time will only be shown in the display.
•
If you press the start button when the instrument is not operating, the timer
starts the temperature control according to the Wait On Timer/ Wait OFF Timer.
•
If you set the Wait OFF Timer while the instrument is in operation, the new
timer will be applied immediately without having to press the Start button again.
•
If you want to change the Wait OFF timer during timer operation, you must turn
off the Wait OFF Timer and reset the process.
4.2.5
Instantaneous Outage Recovery Function
If the power is restored within 30 seconds of an instantaneous power outage, please follow the
following procedure.
Before the Instantaneous
Outage
Operation after Recovery from Instantaneous Outage in Basic
Mode Control
Stopped
Stop
Running
Run
Wait ON Timer is On
The elapsed time of the Wait ON Timer will not be saved, and
will be started with the initial value.
Wait OFF Timer is On
The elapsed time of the Wait OFF Timer will not be saved,
and will be started with the initial value.
Содержание 52412-15
Страница 1: ...User Manual of Forced Convection Oven Model OVF 800 series 52412Mechanicaloven_ver_1 2...
Страница 6: ...52412Mechnicaloven_ver_1 2 6 1 0 Safety...
Страница 16: ...52412Mechnicaloven_ver_1 2 16 2 0 Functional Description...
Страница 23: ...52412Mechnicaloven_ver_1 2 23 3 0 Installation...
Страница 34: ...52412Mechnicaloven_ver_1 2 34 4 0 Operation...
Страница 58: ...52412Mechnicaloven_ver_1 2 58 5 0 Maintenance...
Страница 63: ...52412Mechnicaloven_ver_1 2 63 6 0 Trouble Shooting...
Страница 67: ...52412Mechnicaloven_ver_1 2 67 7 0 Accessories...
Страница 69: ...52412Mechnicaloven_ver_1 2 69 8 0Appendix...