Section 5
Normal Operation
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 83
To display the current fault condition, press the ENTER key while this is displayed.
This display mode is automatically selected by the instrument, should a fault
condition occur. The following fault conditions are recognised by the instrument
:
*ALL CLEAR
*
– No fault condition.
Det. Saturated
– the detector level gain within either the receiver or the signal
processor is too high for the current duct conditions.
Low Det. Level
– detector levels are too low (<3,000).
Mod. Freq. O.R
.
– chopper motor frequency is out of range (<30Hz or >45 Hz).
Reference Fail
.
– no reference signal from the transmitter unit.
Cal. Fact. O.R
.
– after the calibration routine, the calculated Set Cal factor is out of
range – see Fault-Finding.
By pressing the arrow key the previous fault condition can be observed.
5.9.5
Fault Condition
If a fault condition exists, the minutes, hours and days averages will not be updated; see also Data Valid LED
Out.