Section 6
Maintenance
Issue: A Revision: 1 Date: 24/07/19 Ref: 100606
Page: 123
Some instrument components/operation can be verified as follows :
Conduct the procedure for changing the heater cartridge within the transmitter,
(see Replacement of the Heater Element), up to operation 4. Remove one
lead from the heater cartridge and measure the resistance across the two
cartridge terminals. It should be about 3.5
. If it is open circuit the heater
cartridge needs replacing.
Conduct the procedure for adjusting the chopper frequency in the transmitter,
(see
6.5.1
Individual Tests
Heater Cartridge
6.5.1.1
Chopper Motor
6.5.1.2