Oxygen Monitor Operating Manual 

 

Model: RAD-0012 Remote Oxygen Enrichment Safety System 

1. Product Description 

Thank you for selecting the RAD-0012 Remote Oxygen Enrichment Safety System. 
This monitor is designed to detect oxygen enhancement in enclosed spaces and to 
warn occupants of high oxygen levels. High concentrations of oxygen (O2) in 
confined spaces may lead to fire hazards. This monitor has both audible and visual 
alarms which activate when the O2 concentration is higher than the pre-set alarm 
levels. Detection of high levels of O2 will also activate relays that can be used for a 
fan or air-handling system to ventilate the confined space and improve O2 
concentration in the area. 
The RAD-0012 O2 Monitor is cost-effective and has many features including: 
1.

 

Separate SEU (Main Sensor Unit) and RDU (Remote Display Unit) allow you to 
see warning before entering an enclosed area. Up to 3 RDUs can be used. 

2.

 

Large digital LCD display clearly indicates the ambient O2 concentration. 

3.

 

Relay outputs to control ventilation devices. 

4.

 

Audible and visual alarm indications at three separate alarm levels. Ability to 
add strobes for additional indication. 

5.

 

Automatic barometric pressure compensation for high altitude use. 

2. Package Contents & Description   

The RAD-0012 package comprises the following parts: 
SEU (Main Sensor Unit), RDU (Remote Display Unit), Power Supply (Pre-Wired), 
CAT 5 Communication Cable (1 piece), Relay Cables (3 pieces), 
Wall Plug Safety Strap (1 piece), Warning Signs, User Manual (1 piece) 
Mounting Brackets (2 pieces), Screws (13 pieces), Wall Anchors (12 pieces) 
Cable Clips (10 pieces), International Power Adaptor (3 pieces) 
 

SEU (Main Sensor Unit) 

 

 

A. O2 Sensor 

B. LCD display 

C. Power Green LED 

D. Alarm 1 Red LED 

E. Alarm 2 Red LED 

F. Alarm 3 Red LED 

G. Fault Yellow LED 

H. MODE Button 

I. UP Button 

J. DOWN Button 

K. ENTER Button 

L. Alarm Buzzer 

M. Reset Button 

N. 4-20mA Analog 
Output 

O. Battery Backup 
Input 

P. DC Power Supply 

Q. RDU Cable (RJ45) 

R. Relay 3 for AL3 

S. Relay 2 for AL2 

T. Relay 1 for AL1 

U. Strobe Cable (RJ45) 

V. Mount/Panel Holder 

 

 

 

 

RDU (Remote Display Unit)   

 

A. Power Green LED 

B. Alarm 1 Red LED 

C. Alarm 2 Red LED 

D. Alarm 3 Red LED 

E. Fault Yellow LED 

F. LCD display 

G. MODE Button 

H. UP Button 

I. DOWN Button 

J. ENTER Button 

K. Alarm Buzzer 

L. Strobe Cable (RJ45) 

M. Output Cable to 
RDU (RJ45) 

N. Input Cable from 
SEU/RDU (RJ45) 

O. Panel Holder 

3. LCD Display Symbol 

 

 

 

Symbol Meaning 

Description 

 

O2 Concentration 
ppm (Parts Per 
Million) 

Ambient O2 Concentration 

 

Alarm Alarm 

Icon 

DIAG

 Diagnostics 

Test communications between the SEU and 
RDU (see 10.3) 

AL1 

1

st

 O2 Alarm 

Level 

Alarm 1 will trigger when O2 concentration is 
above the first alarm level (default 23.0%). 
AL1 (Red LED 1) and Fault LED will flash, 
buzzer will sound, and Relay 1 will activate. If 
there is strobe, the strobe will flash. This 
status will remain latched. (see 9.1) 

AL2 

2

nd

 O2 Alarm 

Level 

Alarm 2 will trigger when O2 concentration is 
above the second alarm level (default 24.0%). 
AL1, AL2, Fault LED will flash, buzzer will 
sound and relay1 and relay2 will activate. If 
there is strobe, the strobe will flash. This 
status will remain latched. (see 9.1) 

AL3 

3

rd

 O2 Alarm 

Level 

Alarm 3 will trigger when O2 concentration is 
above the third alarm level (default 25.0%). 
AL1, AL2, AL3, Fault LED will flash, buzzer will 
sound and relay1, relay2, relay3 will activate. 
If there is strobe, the strobe will flash. This 
status will remain latched. (see 9.1) 

CALI 

Calibration Icon 

Displays when calibrating the O2 sensor to 
adjust for long-term drift from the actual O2 
concentration. (see 10.7)   

RCFS 

Recover Factory 
Settings 

Reset To factory default settings, removes all 
custom settings. 

Hi 

High 

The O2 concentration is above 25% 

 

Fan Icon 

The fan will appear when alarm levels are 
reached. Any device connected to the relays 
will run. 

 
 
 
 
 

4. SEU (Main Sensor Unit) 

The SEU (Main Sensor Unit) houses the oxygen sensor, connections to the 3 
alarm relays, a strobe (if used) and the connection to the RDU(s). Power for both 
the SEU and RDU are supplied by the SEU. All setup functions and calibration are 
performed from the SEU. 
The SEU should be placed in a room where inert gases like Nitrogen, Argon, and 
others are stored and oxygen depletion can occur. The large LCD displays the 
ambient O2 concentration in real-time. 
There are three separate O2 alarm levels: AL1, AL2 and AL3. All are preset to 
OSHA standards but may be customized (see Section 10). 
When the RAD-0012 Monitor detects an oxygen value above AL1, the Alarm 1 Red 
LED and Yellow Fault LED will blink, Relay 1 will activate, an audible alarm will 
sound, and the Strobe will activate if present. The alarm will continue until the 
RESET button on the SEU is pressed unless the Latch Function is turned off (see 
Section 9.2). 
If the oxygen level continues to increase above AL2, the Alarm 1 Red LED, the 
Alarm 2 Red LED, and Fault Yellow LED will blink, Relay 1 and Relay 2 will 
activate and the audible alarm will sound, The Strobe will activate if present. The 
alarm will continue until the RESET button on the SEU is pressed unless the Latch 
Function is turned off (see Section 9.2). 
If the oxygen level continues to increase above AL3, all 3 Alarm Red LEDs will 
blink, all 3 relays will activate and the audible alarm will sound. The alarm will 
continue until the RESET button on the SEU is pressed unless the Latch Function 
is turned off (see Section 9.2).                               

5. RDU (Remote Display Unit)   

The RDU should be placed outside the enclosed area (typically next to a door) to 
warn users if the oxygen level inside the enclosed area have changed. The RDU is 
controlled and powered by the SEU. A strobe may be attached to the RDU. 
All visual and buzzer alarms on the SEU are duplicated on the RD. The SEU only 
has the “DIAG” function to test the communication between the SEU and RDU 
(see 10.3). All other functions or settings must be changed on the SEU. 

6. Strobes   

Strobes are add-on visible alarms. One strobe can be connected to the RDU and 
one to the SEU. If the oxygen level goes below Alarm Level1, the strobes will flash. 
The frequency of the flash cannot be changed. 

7. Power   

The RAD-0012 comes pre-wired with a 12V power supply that plugs in to a 120-
240VAC wall outlet. The 12V power supply can be removed and 24VDC can be 
wired directly to the SEU through the terminal block. 24VAC must be converted to 
24VDC for the monitor to operate.   

8. Installation 

1.

 

Choose a suitable location near a wall outlet to install the SEU. Fix the panel 
holder on the wall with the four screws provided. 

2.

 

Put the SEU on the panel holder, making sure that they are connected tightly. 

3.

 

Fix the second panel holder in a suitable location outside the monitored space 
at eye level. Place the RDU onto the panel holder. Display warning signs next to 
the RDU so they are not hidden when the door is open. 

4.

 

Route the included CAT5 cable between the SEU and RDU. CAT5 cable can be 
run through the wall/conduit or fixed to the wall using cable clips. Plug the CAT5 
cable into the designated ports on both units. 2 additional RDUs can be 
connected to the first RDU as long as the total cable length between the farthest 
RDU and SEU is less than 300ft. (91m). 

5.

 

The RAD-0012 has 3 relay outputs connected to the programmed alarm settings. 
All relays are normally open/closed dry contacts. Any of the relays can be used 
to control an external device (fan, HVAC system, etc.) or can be wired to the fire 
alarm panel directly. The relays will trigger when the CO2 concentration exceeds 
the programmed alarm level. 

6.

 

When the power has been connected, The SEU and the RDU will perform a self-
check, then begin to work. If the cable between the SEU & RDU is not securely 
connected, the yellow fault LED on the SEU will blink after startup. If cable is 
inserted into the wrong port on RDU, after about one minute, “Er7” will flash on 
the RDU display. Securely plug the cable into the correct port on RDU for the 
unit to function normally. 

 

7.

 

To test the system, use the DIAG function. The five LED’s will blink and the 
buzzer will sound on both the SEU and RDU. Then both LCD screens will show 
the same information. This verifies that alarm is ready. 

Installation Example: 

 

9. Advanced Management Settings 

These settings use a non-obvious key combination to prevent casual users from 
changing alarm settings. In most cases, the default settings are recommended. 

9.1 Alarm Level Lock / Unlock:

 If the alarms are locked, an employee cannot 

change the alarm levels. Factory default is locked. You must unlock the alarm 
levels before you can change them (see Section 10). 

 

1. Press Up & Enter buttons simultaneously for 10 seconds until BAro is displayed. 
2. Press Mode button until Al n is displayed. 
3. Press Up/Down arrow buttons to switch between Lock and Unlock. 
4. Press Enter button to save and quit the advanced mode. 

 
9.2 Latch Function On / Off:

 If the Latch function is on, alarms will continue until 

the reset button is pressed even if the O2 level returns to normal. Factory default is 
on.

 

 

1. Press Up & Enter buttons simultaneously for 10 seconds until BAro is displayed. 
2. Press Mode button until LAt is displayed. 
3. Press Up/Down arrow buttons to switch between On and Off. 
4. Press Enter button to save and quit the advanced mode 

9.3 Barometric Compensation On / Off:

 Turns on automatic compensation for 

barometric pressure / altitude. Factory default is on. 

 

1. Press Up & Enter buttons simultaneously for 10 seconds until BAro is displayed. 
2. Press Mode button until BAro is displayed. 
3. Press Up / Down arrow buttons to switch On or Off. 
4.Press Enter button to save and quit the advanced mode 

 

 