CME Medical BodyGuard 323 Скачать руководство пользователя страница 1

 

 
 
 

 
 

 

 

 
 
 

 
 

 

BodyGuard 323

 

Infusion Pump System 

Operator Manual 

 

 

 

0344 

                    

Manufacturer:                                                                      

Caesarea Medical Electronics Ltd. 

                                 

         

 

European address: 
Staufenburgstr. 23 Lichtenstein  P.O.B. 1248  
Lichtenstein 72805 Germany                                
International Toll Free: +800-323-575-00                                                                 

www.cme-infusion.com

 

Contact e-mail: [email protected] 

 

                                                                                           

                           

 

EDITION July 2009 

 

Ref.: 100-090X  

 

Содержание BodyGuard 323

Страница 1: ...anufacturer Caesarea Medical Electronics Ltd European address Staufenburgstr 23 Lichtenstein P O B 1248 Lichtenstein 72805 Germany International Toll Free 800 323 575 00 www cme infusion com Contact e mail Sales cme infusion com EDITION July 2009 Ref 100 090X ...

Страница 2: ...ions and follow all recommendations Using the pump with air in line detector off may cause an embolism resulting death or paralysis Switching the air detector off is not recommended and should be used only when absolutely necessary Please contact your technician in order to enable the air sensor if disabled with no need U S A federal law restricts this device to sale by or on the order of a physic...

Страница 3: ...sion Alarm Test 22 Door Open Alarm Test 23 Battery Test 24 Charger Indicator Test 24 Dedicated BodyGuard Administration Sets 25 The BodyValve 25 Loading the Administration Set 29 Priming the Line 30 Prime In Continuous Intermittent TPN and 25 Steps Modes 30 Prime In PCA Mode 31 Prime Operation After Air In Line Alarm 32 Manual Prime Operation 32 Operation steps 33 Pump Mounted on IV Pole 33 Portab...

Страница 4: ... Lock 65 Info Mode 67 Review Set up 69 Change Set up 70 5 Alarm Conditions and Troubleshooting 72 Alarm Condition 72 Troubleshooting 73 6 Specifications 75 7 Service and Maintenance 77 Cleaning 77 Storage 77 Battery Operation 78 LIMITED WARRANTY 79 ...

Страница 5: ... Analgesia Continuous delivery plus or programmed boluses Bolus Volume 0 25 5 ml Clinician Bolus Volume 0 1 100ml Bolus Rate 0 1 1200 ml h Operator can design a specific protocol for drug delivery which contains up to 25 steps Infusion Rate 0 1 to 1200 ml hr Volume up to 9999ml for each step Accumulated Volume for all steps up to 10 liters Highly accurate fluid delivery Operated with a custom safe...

Страница 6: ...s programming choices and instructions 2 Up Arrow Scrolls up through options 3 START OK Starts infusion Confirms selection and setting 4 STOP NO Stops infusion Silences an alarm condition Pause priming Zeroes the displayed value during programming Erases the last digit during programming Returns to previous screen 10 1 3 2 5 6 4 8 7 9 ...

Страница 7: ...on until the Self Test screen appears Turn the system OFF by pressing and holding the button until the graph is black and a beep is generated 7 Operation LED Green Indicator Lights during system Self Test An intermittent green light indicates infusion delivery Red Indicator Indicates an alarm state with a continuous red light Lights when the pump is in a stand by mode during programming 8 INFO Sup...

Страница 8: ... level can be adjusted to suit patient needs 3 Keyway Guides the IV line 4 Pressing Plate Connected to the door by two springs 5 Pump Door Covers the pressing plate 6 Flow Direction Shows direction of pump operation 7 Anti Stretch locator Optional Locates the tube in the correct length from the key 8 Air sensor Front Ultrasonic air detector which is mounted on the front housing 9 Air sensor Door M...

Страница 9: ...1 Introduction BodyGuard Operator Manual 9 The Charger Area Function 1 Charger Display Displays major pump messages and operation values 2 Release Handle Push to release the pump from the Charger 2 1 ...

Страница 10: ...ume of air that passes through the IV line by accumulating the volume of individual bubbles over any 15 minutes period This limit is not configurable Although an individual bubble may not exceed the pre programmed threshold if the cumulative volume of smaller bubbles exceeds 1ml over a period of 15 minutes an air in line alarm is initiated This accumulation feature is particularly useful with infu...

Страница 11: ...EC 60601 2 24 standard Trumpet and Flow Rate With the BodyGuard 323 as with all infusion systems the action of the pumping mechanism and variations cause short term fluctuations in rate accuracy The following curves show typical performance of the system in two ways 1 The accuracy of fluid delivery over various time periods is measured trumpet curves 2 The delay in onset of fluid flow when infusio...

Страница 12: ...ndows may be of interest when certain drugs are being administered Short term fluctuations in rate accuracy may have clinical impact depending on the shelf life of the drug being infused and the degree of inter vascular integration The clinical effect cannot be determined from the trumpet curves alone ...

Страница 13: ...1 Introduction BodyGuard Operator Manual 13 ...

Страница 14: ...1 Introduction BodyGuard Operator Manual 14 ...

Страница 15: ...uctions 0344 CE mark indicates conformance to Medical Device Directive 93 42 EEC Do not dispose of in municipal waste Symbol indicates separate collection for electrical and electronic equipment WEEE Directive 2002 96 EEC NOTE Does not apply to the battery Do not dispose of battery in municipal waste Symbol indicates separate collection for battery is required The use of single use disposable comp...

Страница 16: ... enteral in close proximity to nerves and into an intraoperative site soft tissue body cavity surgical wound site The system is intended for patients who require maintenance medications analgesics PCA therapy parenteral and enteral nutrition fluids chemotherapeutic agents and general fluids therapy in hospital and home care environments Warnings To avoid possible personal injury or loss of life ob...

Страница 17: ...tments maintenance or repair of the uncovered pump or charger while connected to the power should be avoided The BodyGuard Infusion System should be operated within a temperature range of 18 C 59 F to 45 C 113 F and up to 85 humidity Operating the pump at temperatures and or humidity other than within that range may affect accuracy Unsafe operation may result from using improper accessories Use on...

Страница 18: ...y in a depleted state for a long period of time may damage the battery Connect the pump to the mains via the Charger whenever possible to recharge the battery Do not store the pump with the battery fully depleted Xylene Acetone or similar solvents could cause damage to components Do not clean the pump with these chemicals Clean solution spills on the pump immediately Use a damp cloth or sponge A m...

Страница 19: ... following items BodyGuard Infusion Pump Charging Cradle with Power Cord or Battery Charging Cable Rechargeable Battery Pack Operation Manual PCA bolus cable optional 9V Alkaline Battery compartment optional Disposable Battery Pack optional External charger optional Extended life battery Li Ion Polymer 3600 mAh optional External wall charger Optional If any items are missing or damaged contact you...

Страница 20: ...rger unit to AC power and verify that the AC indicator is lit 2 Put the pump into the charger top first until it clicks into place 3 The battery is charging when the red battery LED on the front of the charger is lit The battery is fully charged when the battery LED symbol turns green If the led does not light or does not turn green after more than 8 hours pls replace battery 4 Remove the pump fro...

Страница 21: ... NOTE For instructions on how to load and prime an administration set refer to sections Loading the Administration Set Priming the Line 3 Press until the Self Test screen appears If air sensor is off press to confirm 4 The program selected and its status are displayed During the Self Test a long beep will be heard This procedure verifies that the pump s acoustic and visual features are working pro...

Страница 22: ...rate at 500ml hr using the Numeric keypad and press to confirm setting 5 Set Volume at 20 ml and press to start operation 6 Insert air into the IV line to allow an air bubble greater than 10 mm to enter into the administration set 7 Air in line alarm will be activated as soon as the air bubble enters the set segment located behind the pump door Press to turn the alarm off Down Occlusion Alarm Test...

Страница 23: ... set flow rate and pressure threshold the Down Occlusion alarm will be activated Press to mute the alarm Door Open Alarm Test 1 Insert a BodySet in the pump and close the door Press the door until a click is heard NOTE For instructions on how to load and prime an administration set refer to sections Loading the Administration Set Priming the Line 2 Press until the Self Test screen appears If air s...

Страница 24: ... instructions Set the infusion rate at 1200 ml hr using the Numeric keypad and press 6 Set Volume at 9999 ml and press to start operation 7 Wait untill pump displays low battery message 8 Press to turn the pump off 9 Connect the pump to a charger 10 Verify the red icon of battery is lit on the charger 11 Verify the red battery turns green after a minimum of 2 hours WARNING Do not let the pump oper...

Страница 25: ...nical malfunction Preventing pump operation if the set has been loaded incorrectly Warning Replace the administration set every 72 hours to lessen the incidence of bacteria formation Warning Disposables must be compatible with the medicine delivered Warning Patient or operator injury may result if package is opened or damaged or if damaged components are used Visually inspect contents and package ...

Страница 26: ...le Connects the administration set to solution bag 3 Slide Clamp Can be used to occlude the tube 4 Key Locates the tube in the correct flow direction 5 Administration Tubing PVC 3 x 4 1 mm Transfers the solution 6 Y Connector Used to add drugs by syringe 7 BodyValve With Cap Anti free flow valve 8 Valve Cap Keeps the set sterile ...

Страница 27: ...cts the administration set to solution bag 2 Drip Chamber Displays infused drops from the administration bag to the patient 3 Slide Clamp Can be used to occlude the tube 4 Key Locates the tube in the correct flow direction 5 Administration Tubing 3 x 4 1 mm Transfers the solution 6 Y Connector Used to add drugs by syringe 7 BodyValve With Cap Anti free flow valve ...

Страница 28: ...e luer lock Connects the administration set to the patient access device 6 Male luer lock Cap Keeps the set sterile 7 Administration Tubing 3x 4 1 mm Transfers the solution pumping segement 8 Slide clamp Can be used to occlude the tube 9 Prime valve Anti free flow valve with manual priming button 10 Key Locates the tube in the correct flow direction 11 One way valve Prevents free flow and back flo...

Страница 29: ...ng the key into the keyway as shown by the arrow in diagram 1 Insert the tubing from left to right with the shorter end connected to the medication bag on the left and avoid stretching or pulling the tubing Check that the key located on the tube is located in its correct position in the tubing guide NOTE The IV key can only be fitted into the key space one way If you are having trouble fitting it ...

Страница 30: ...the priming procedure in PCA mode is different from all other programs For detailed instructions refer to PCA Operation section 1 Turn on the BodyGuard Pump by pressing until the Self Test screen appears If air sensor is off press to confirm 2 When the display screen shows the existing programmed rate press once Warning Ensure the set is not connected to the patient 3 Enter the Volume of fluid req...

Страница 31: ... to prime the set and press to begin priming procedure You may stop priming at any time by pressing 5 The display screen will show a graph indicating the prime operation 6 Once the priming is completed the display will change back to the previous screen the pump was in before priming If more priming is needed repeat procedure Make sure that all air is removed from the IV line before connecting to ...

Страница 32: ...line 7 Connect IV line to the patient and press to resume program Manual Prime Operation 1 In any case before operation or after air in line alarm a prime operation is needed the user can prime the set manually by using the manual priming valve if assembled in the administration set in use at that moment 2 Disconnect IV tubing from patient 3 Hold the tubing in a vertical position to maintain gravi...

Страница 33: ...l appear on the display followed by the last rate setting 7 Prime the administration set Warning Make sure patient is not connected to the pump during priming 8 Set the required program and continue as described for every program setting Warning Verify that infusion is proceeding normally before leaving the pump unattended Portable Connection 1 Connect the Bodyset MicroSet to the medication bag 2 ...

Страница 34: ...per up taper down pattern The parameters to be set are Total Volume Time Up Time Down and Total Time The pump will calculate the ramping infusion rates PCA Program Used for PCA applications Combines basal rate and pre programmed boluses Rate 0 0 to 100 ml hr in 0 1 ml increments Volume 0 1 to 1000 ml Bolus Volume 0 to 100 ml Bolus Rate 0 1 to 100 ml hr in 0 1 ml increments and 100 to 1200 ml hr in...

Страница 35: ...rd and press 6 Scroll down to Select Program option and press 7 Scroll to the required program and press 8 Press to restart the pump NOTE The selected program will remain in the pump s memory until changed by the operator Changing Current Program Short Way 1 Turn on the BodyGuard Pump by pressing until the Self Test screen appears If air sensor is OFF press to confirm 2 Once the Self Test is compl...

Страница 36: ...lock out mode parameters can be viewed but cannot be changed NOTE If pump is not in Continuous Program refer to changing program section for instructions NOTE Whenever you decide to set a new program follow instructions at the bottom of screen NOTE The screen will be blank upon initial setup When returning to a program the screen will display the last set program 3 Rate Screen Do one of the follow...

Страница 37: ...ired rate during operation and press within five seconds The pump will briefly alarm and the infusion will resume at the adjusted rate NOTE If the rate change is not confirmed within ten seconds the pump will continue operation at the original rate and the display will return to the previous setting 2 To stop the infusion press To change the rate while in Stop mode 1 Press any number on the keypad...

Страница 38: ...irm the existing parameters Enter the desired volume using the numeric keypad and press NOTE The screen will be blank upon initial setup 9 Infusion Time Screen Enter the desired total time using the numeric keypad for example enter 0 2 0 0 for two hours and press NOTE First two digits represent hours and last two digits represent minutes 10 Rate Screen The pump calculates the rate based on the tim...

Страница 39: ... mute or to continue Resume or Repeat Infusion Unlocked When the pump is restarted in Continuous mode and is not locked the pump will allow the user to confirm each of the parameters of the infusion The Volume will be adjusted for the amount infused prior to turning off the pump To reset the volume use the numeric keypad to enter new volume Locked 1 When the pump is turned OFF in the middle of an ...

Страница 40: ...press 4 Pump will beep once and resume the Primary Infusion when the Secondary Infusion is completed NOTE User can stop the Secondary Infusion at any time by pressing and turning the pump off When the pump is turned back on the pump will resume the Primary Infusion NOTE User can titrate the Secondary Infusion with the same method that is used for a primary continuous infusion P I G Y P I G Y Secon...

Страница 41: ...ng and unlocking the pump refer to lock mode section If pump is unocked the Bag Volume screen will display after 2 seconds NOTE When pump is in Program mode all parameters can be adjusted When pump is in lock out mode parameters can be viewed but cannot be changed NOTE If pump is not already in Intermittent Program refer to Changing Program section for instructions NOTE Whenever you decide to set ...

Страница 42: ...art Date For further information regarding the Delay Option refer to Delay Program section 9 Press to Start Infusion 10 During program operation the pump displays the rate in which the program is operating NOTE If one of the values is changed during the programming mode the pump will not recall the new value unless changes are confirmed by pressing NOTE If prime procedure is activated during data ...

Страница 43: ...uring dose delivery and not during KVO between doses NOTE If the rate change is not confirmed within ten seconds the pump will continue operation at the original rate and the display will return to the previous setting 2 To stop the infusion press To change the rate while in Stop mode 1 Press any number on the keypad twice to return to the rate screen 2 Adjust the rate and press 3 Adjust all other...

Страница 44: ...ting parameters Enter the Bag Volume and press 9 Dose Volume Screen Do one of the following Press to confirm the existing parameters Enter the Dose Volume and press 10 Dose Time Screen Do one of the following Enter the Dose Time and press To enter 30 minutes press 0 0 3 0 11 Interval Time Screen Do one of the following Press to confirm the existing parameters Enter the time between the start of ea...

Страница 45: ...termittent infusion is unlocked and stopped and the pump is re started the user has the following options 1 Resume Resume the original infusion exactly where it left off 2 New Bag Resume the original infusion with a new bag 3 Repeat Repeat or re program the infusion 1 Intermittent Restart Screen Press or to scroll to the appropriate selection and press NOTE If pump is Locked the user will only hav...

Страница 46: ...ess 6 Scroll to Delay Before Start option and press 6 Turn delay option ON by pressing or 7 Press and the screen displays Restart Pump 8 Press to continue Setting Delay Before Start After entering the program data the pump enables delay setting when Delay Before Start is turned ON NOTE The delay time is set according to a 24 hour clock This means that hours are viewed as follows 10 00 11 00 12 00 ...

Страница 47: ... the user can restart the infusion within 1 hour and 20 minutes of the interval time to receive the remainder of the dose In this case the pump will still resume the following dose at the scheduled time If the user restarts the infusion after one third of the interval time has elapsed then the user will not be receive the remainder of the dose The pump will resume in the middle of the interval and...

Страница 48: ... absolutely necessary 2 Pump displays program name and status The status is either Lock out or program for instruction regarding locking and unlocking the pump refer to locking section When the pump is unlocked press to continue NOTE When pump is in Program mode all parameters can be adjusted When pump is in lock out mode parameters can be viewed but cannot be changed NOTE If pump is not already i...

Страница 49: ...es which step is currently running and the rate NOTE If one of the values is changed during the programming mode the pump will not recall the new value unless changes are confirmed by pressing If prime procedure is activated during data setting the pump will return to the original previous screen once priming is completed NOTE Whenever required to stop infusion press Stop displays on the screen St...

Страница 50: ...riod in addition to the 5 ml KVO volume 1 Turn the pump on by pressing and holding the until the Self Test screen appears Warning If air sensor is off please contact your technician in order to enable the air sensor if disabled Using the pump with air in line detector off may cause an embolism resulting death or paralysis Switching the air detector off is not recommended and should be used only wh...

Страница 51: ...the desired value from 00 00 to 04 15 hours Press 8 KVO Rate Screen Do one of the following Press to confirm the existing parameters Enter the desired Interval Time and press 9 Accumulation Screen The pump calculates the set data and displays the accumulation screen with the maximal rate for the program to be infused during flat period Do one of the following Press to confirm parameters and start ...

Страница 52: ...urn to the original previous screen once priming is completed NOTE Whenever required to stop infusion press Stop displays on the screen Stop mode is limited for two minutes After two minutes an audible alarm is sounded Press again to mute the alarm NOTE During programming you may backspace by using Pressing once clears the last entered digit Pressing again clears the next number Pressing when the ...

Страница 53: ...air in line detector off may cause an embolism resulting death or paralysis Switching the air detector off is not recommended and should be used only when absolutely necessary 2 Pump displays program name and status The status is either PROGRAM or LOCKED for instruction regarding locking and unlocking the pump refer to Unlocking PCA section Press If the pump is unlocked the PCA Menu will display I...

Страница 54: ...een will show a graph indicating the prime operation 5 Once the priming is completed the display will change back to the previous screen the pump was in before priming If more priming is needed repeat procedure Program Infusion 1 PCA Menu Screen Select Program Infusion and press 2 New Patient Screen If programming a protocol for a new patient select New Patient by pressing If the protocol is not f...

Страница 55: ...itration that can be used in the current protocol and press Rate can be ml h mg h or mcg h depending on program selected 7 Bolus Screen Do one of the following Bolus can be ml mg or mcg depending on the program selected Press to confirm the existing parameters Enter the desired bolus dose and press 8 Bolus Titration Limit Screen Enter the maximum bolus dose that can be used in the current protocol...

Страница 56: ...us step for the hour limit Do one of the following Limit can be ml mg or µg depending on the program selected Press to confirm the existing parameters Enter the desired hours for the limit and press 12 Bag Volume Screen Do one of the following Press to confirm the existing parameters Enter the desired bolus volume and press 13 Locking Screen The pump automatically locks before a PCA infusion progr...

Страница 57: ...be delivered over a longer period in Change Set Up using the Technician Access code Interpreting the Display Screen During PCA Infusion Infusion Mode The first line of the display indicates that the pump is in the PCA infusion mode Infusion Rate The second line of the display shows the current infusion rate Depending on whether the user programmed the current protocol in mls or mgs this will displ...

Страница 58: ...olus before starting the PCA infusion perform the following steps 1 Selecting the Prime option from the PCA menu the screen will show the priming screen Select Clinician bolus option 2 Enter Level 2 Technician Code and press 3 Enter the volume of the Clinician Bolus and press to begin priming procedure You may stop the bolus at any time by pressing The Volume for the Bolus will be in ml mg or µg d...

Страница 59: ...nician Code and press 3 Enter the volume of the Clinician Bolus and press to begin priming procedure You may stop the bolus at any time by pressing The Volume for the Bolus will be in ml mg or µg depending on the settings of the existing protocol entered for the PCA infusion 4 The display screen will show a graph indicating the Bolus is being infused The graphical bar displays the amount of the bo...

Страница 60: ...eft off previously New Bag Restarts infusion exactly where it left off previously and updates bag volume to be equal to the bag volume originally programmed Menu Brings user back to the PCA Menu When asked for a code enter the Unlock L2 Code NOTE If using an infusion set with drip chamber change to a new infusion bag but make sure that the drip chamber is half filled and there is no air in the adm...

Страница 61: ...een showing the currently programmed Bolus in ml or mg Use the keypad to enter the new Bolus and confirm the new Bolus by entering the Level 1 Code The Bolus can be increased up to the Maximum Bolus Dose programmed into the specific infusion Changing Lock Out Time During The Infusion 1 Stop the infusion by pressing 2 Then press and hold until the screen appears allowing you to access the PCA Menu ...

Страница 62: ...livered During the Last 24 Hours Hour by Hour NOTE Press to resume infusion when you return to the STOP screen Events Log To access the history of events you need to access Change Set Up using the Level 1 Technician Access Code The events history carries the last 500 user events i e changes to programs pump started stopped etc and the status of the pump i e battery charge pressure levels etc It is...

Страница 63: ... through the information shown here to check what the pump was set up for whether it was operating on battery or mains whether it was locked out and what the pressure settings and readings were at that time Rate 1ml h Volume Left 99 4 Infused 0 1ml PCA Program Battery Operation Normal Level Battery Pressure Level Taken Pump Unlocked Vol Cal 540 Pressure Setting High Pressure DELTA 15 Pressure CAP ...

Страница 64: ...ault and press 6 Press or to adjust the level high normal low 7 Press to store the parameter in memory NOTE The following pressure parameters can be set Low 7 psi Normal 14 psi High 21 psi The selected pressure level will remain in memory until it is changed The actual pressure can be viewed at any time by scrolling with the info button during operation After a down stream occlusion infusion will ...

Страница 65: ...ure enables the operator to set a maximal rate for the pump to accept The program rate then can be adjusted only up to a preset limit How To Lock A Keypad Locking To Lock Keypad Press and hold until the entire graph is black and a beep is heard The beep indicates that the locking program is turned ON NOTE The keys are locked in memory until they are unlocked or until entering Change setup mode To ...

Страница 66: ... STOP or setting mode 2 Scroll to Change Set up option Press 3 Scroll to More and press 4 Enter Change Set up Level 1 code using the numeric keyboard and press 5 Scroll to Maximal Rate option and press Use numeric keys to set desired rate and press Bolus Rate Maximal Rate KVO Rate Volume Infused Review Set up Change Set up Buzzer Level More Change Set up Code Technician XXX Program Lock OFF Adjust...

Страница 67: ...t 25 C Three Continuous Intermittent TPN Tubing Temperature P C A Boluses Attempted 3 Given 1 Delay 60Min Three PCA Boluses History for current program Time Left 1 00 Four Continuous Intermittent Time remaining for the end of the program Time 0 6 30 12 Four TPN Program Status in a graphic way P A Rate 50 ml h Bolus 5 ML Lock Time 60Min 4H Limit 100 ml Four PCA Data of current protocol Time 12 55 D...

Страница 68: ...w instructions in the Change setup section To view current pump settings select Review setup press Start OK and follow instruction in Review setup section NOTE If no selection is made within ten seconds after pressing the screen will exit the information mode and display the previous screen To exit the information mode after making a selection in Review Set up press once to return to the original ...

Страница 69: ...t Program Current Selected Program 9 Priming Rate Current Priming Rate Setting 10 Maximal Bolus Volume Maximal Bolus Volume 11 Bolus Rate Current Bolus Rate Setting 12 Maximal Rate Current Maximal Rate Setting 13 KVO Rate Current KVO Rate Setting 14 Flow Control Flow Control 15 Air Bubble Size Current Air Bubble Size Setting 16 Store Last Program Option Currently ON or OFF 17 Operation LED Option ...

Страница 70: ...l 2 Technician Code using the numeric keyboard and press 3 Press or to change from OFF to ON and press to approve To Adjust Other Parameters 1 Scroll to More and press 2 Enter Level 1 Technician Code using the numeric keyboard and press 3 Press or to select an option and press 4 Adjust the parameter by pressing or or by using the numeric keypad Buzzer Level More Change Set up Code Technician XXX B...

Страница 71: ...he max rate from 0 1 1200 ml hr 11 KVO rate Set the KVO rate from 0 1 5 0 ml hr 12 Flow Control 0 20 drops per ml 13 Air bubble size Set detectable size from OFF to 1 0 ml 14 Store last program Turn on to clear all program parameters upon system shut off Turn off to save all program parameters upon system shut off 15 Operation LED Turn operation LED ON OFF 16 Time option Turn time option ON OFF Wh...

Страница 72: ...larm Condition When the infusion pump detects a problem four things occur The infusion stops An audible alarm is activated A message appears on the display screen indicating the cause of the alarm and The LED indicator will change from green to red NOTE Pressing during an alarm mutes the alarm for two minutes ...

Страница 73: ...Change the Cannula Re load the IV line Refer to Setting Down Occlusion section Pump Unattended An alarm is activated Two minutes has elapsed without a button press during programming Press START OK to resume Door Open Infusion stops and an alarm is activated The door of the pump was not closed prior to operation The door has accidentally opened during operation Close the door of the pump Send the ...

Страница 74: ...oaded in the pump Reload the set into the pump as per the instructions Change the set to a BodySet Microset and load it according to instructions Lock Out Setting cannot be changed Lock Out mode is turned on No action required If changing program is required unlock the pump and refer to lock section for further instructions Lock mode Setting keys do not function Lock Out mode is turned on No actio...

Страница 75: ...hr Air Sensor Ultrasonic adjustable air bubble size Maximum Pressure 21 psi Adjustable High 21 Normal 14 Low 7 Power Supply 110 240 VAC 50 60 Hz 0 3A max Battery Rechargeable Li Ion 7 2V 1800mAh Typical Battery Operation at 125 ml hr 20 hours Rechargeable Battery Charging Automatic when clicked into the Charger that is connected to an AC power source Six hours needed to charge a fully depleted bat...

Страница 76: ... 2003 CE marked In accordance with the Medical Devices Directive 93 42 EEC EMC The BodyGuard Infusion System is designed to be in compliance with EN 60601 1 safety and IEC 601 1 2 EMC Environmental Specifications Non Operating Conditions Transportation and Storage Temperature 25 o C to 50 o C 13 o F to 122 o F Humidity 5 to 100 R H non condensing Air pressure 48kPa to 110kPa Operating Conditions T...

Страница 77: ... clean the pump or charger with chemicals such as Xylene Acetone or similar solvents These chemicals can cause damage to plastic components and paint Use a lint free cloth dampened with warm water and a mild detergent or disinfectant Caution Do not soak or immerse any part of the pump or charger in water Storage If the pump is to be stored for an extended period it should be cleaned and the batter...

Страница 78: ...k up Note When the pump is not in use click the pump into the Charger and plug the system into an AC wall outlet if possible to charge battery Note After the End Battery signal has been activated or following long periods of storage wait 2 minutes after the pump has been connected to an AC power supply before operating Caution Leaving the battery in an uncharged state for a long period of time may...

Страница 79: ...cement Shipping costs are the purchaser s responsibility Material returned to CME or its designated representative should be properly packaged using CME shipping cartons and inserts Inadequate packaging may result in severe pump damage This Limited Warranty shall not apply to defects or damage caused wholly or in part by negligence spilt fluids dropping of the pump misuse abuse improper installati...

Страница 80: ... 58 62 67 71 I Info 7 16 42 45 49 51 53 58 61 63 64 65 67 68 69 71 Infused Volume 67 68 75 Intended Use 16 Intermittent Program 34 41 43 IV pole 5 17 33 K Keep Vein Open Mode 34 Keypad Locking 8 29 Keyway 8 29 KVO rate 34 42 45 49 51 70 72 76 L Lock Mode 36 41 65 73 74 Lock Out 36 41 48 50 61 65 70 71 74 75 M Maximal Rate Locking 65 66 Missing Key 74 75 N Numeric Keys 7 22 24 35 37 39 43 44 46 47 ...