![CLB W1600 Скачать руководство пользователя страница 12](http://html.mh-extra.com/html/clb/w1600/w1600_manual_2619901012.webp)
I-FORM
Subject:
W1600 WAMM
Reference: U1283_RQ_2210
Status:
Pending
Version:
9
Approver: XX
Filename: U1283_RQ_2210_IFORM_W1600.doc
page 12 of 29
1.4.5
Actuator Specifications
1.4.5.1
Front actuators
Image 6: W1600 front actuators
Index
Function
Description
1
Button 1
Switch to make a Nurse Call
2
Button 2
Default: switch for presence
3
Button 3
Default: switch to make an assistance call
4
Button 4
Default: none
5
Button 5
Default: none
6
Light sensor
Ambient light measurement for brightness adjustment
7
Microphone
Microphone for speech or AM2
8
Identify
Activated by holding a magnet at position
≤
2 seconds:
Network presence (Identify) notification of W1600 to the
system. 1 beep can be heard when magnet is positioned
correctly.
≥
5 seconds:
Safe shutdown (manual):
-
Position magnet until 1 short beep is heard and
hold at this position for ~5 sec
2 short beeps will be heard, as indication that
the W1600 is shutting down. The magnet can
now be removed.
-
Wait
until 1 long beep will be heard (~5 sec.).
The W1600 can now safely be unplugged from
power.
1
2
3
4
5
7
6
8
8